Rapid Switch from Intra-Aortic Balloon Pumping to Percutaneous Cardiopulmonary Support Using Perclose ProGlide
We present a case of a patient who needed rapid switch from intra-aortic balloon pumping (IABP) to percutaneous cardiopulmonary support (PCPS)/venoarterial extracorporeal membrane oxygenation. It is difficult to switch from IABP to PCPS, because 0.035-inch guidewires cannot pass the IABP guidewire l...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2015-01-01
|
Series: | Case Reports in Cardiology |
Online Access: | http://dx.doi.org/10.1155/2015/407059 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We present a case of a patient who needed rapid switch from intra-aortic balloon pumping (IABP) to percutaneous cardiopulmonary support (PCPS)/venoarterial extracorporeal membrane oxygenation. It is difficult to switch from IABP to PCPS, because 0.035-inch guidewires cannot pass the IABP guidewire lumen (0.025-inch compatible), and the IABP sheath needs to be removed together with the IABP catheter. First, a 0.025-inch guidewire was inserted into the IABP wire lumen, and then the IABP catheter together with the 8 Fr IABP sheath was removed, leaving the 0.025-inch guidewire in place. We used the Perclose ProGlide for safe and rapid exchange of the 0.025-inch guidewire for a 0.035-inch guidewire. This allowed insertion of a PCPS cannula and the prompt initiation of PCPS. |
---|---|
ISSN: | 2090-6404 2090-6412 |