Liquid Membrane pH Electrode Based on a Synthetic Proton Carrier

A new synthetic H+-selective ion-carrier suitable as a component for liquid membrane electrodes is described. Electrodes based on PVC membranes make possible the measurement of physiological H+ ion activities in the presence of typical extra- or intracellular backgrounds of the interfering ions Na+...