Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner

The aim of this study is to simulate GE Discovery 690 VCT positron emission tomography/computed tomography (PET/CT) scanner using Geant4 Application for Tomographic Emission (GATE) simulation package (version 8). Then, we assess the performance of scanner by comparing measured and simulated paramete...

Full description

Saved in:
Bibliographic Details
Main Authors: Elham Kashian, Hadi Ahangari, Vahab Dehlaghi, Karim Khoshgard, Pardis Ghafarian, Raheb Ghorbani
Format: Article
Language:English
Published: Thieme Medical and Scientific Publishers Pvt. Ltd. 2020-10-01
Series:World Journal of Nuclear Medicine
Subjects:
Online Access:http://www.thieme-connect.de/DOI/DOI?10.4103/wjnm.WJNM_4_20
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850228295400947712
author Elham Kashian
Hadi Ahangari
Vahab Dehlaghi
Karim Khoshgard
Pardis Ghafarian
Raheb Ghorbani
author_facet Elham Kashian
Hadi Ahangari
Vahab Dehlaghi
Karim Khoshgard
Pardis Ghafarian
Raheb Ghorbani
author_sort Elham Kashian
collection DOAJ
description The aim of this study is to simulate GE Discovery 690 VCT positron emission tomography/computed tomography (PET/CT) scanner using Geant4 Application for Tomographic Emission (GATE) simulation package (version 8). Then, we assess the performance of scanner by comparing measured and simulated parameter results. Detection system and geometry of PET scanner that consists of 13,824 LYSO crystals designed in 256 blocks and 24 ring detectors were modeled. In order to achieve a precise model, we verified scanner model. Validation was based on a comparison between simulation data and experimental results obtained with this scanner in the same situation. Parameters used for validation were sensitivity, spatial resolution, and contrast. Image quality assessment was done based on comparing the contrast recovery coefficient (CRC) of simulated and measured images. The findings demonstrate that the mean difference between simulated and measured sensitivity is <7%. The simulated spatial resolution agreed to within <5.5% of the measured values. Contrast results had a slight divergence within the range below 4%. The image quality validation study demonstrated an acceptable agreement in CRC for 8:1 and 2:1 source-to-background activity ratio. Validated performance parameters showed good agreement between experimental data and simulated results and demonstrated that GATE is a valid simulation tool for simulating this scanner model. The simulated model of this scanner can be used for future studies regarding optimization of image reconstruction algorithms and emission acquisition protocols.
format Article
id doaj-art-f9f395eb5bdf415d99d5f6765eecdfda
institution OA Journals
issn 1450-1147
1607-3312
language English
publishDate 2020-10-01
publisher Thieme Medical and Scientific Publishers Pvt. Ltd.
record_format Article
series World Journal of Nuclear Medicine
spelling doaj-art-f9f395eb5bdf415d99d5f6765eecdfda2025-08-20T02:04:34ZengThieme Medical and Scientific Publishers Pvt. Ltd.World Journal of Nuclear Medicine1450-11471607-33122020-10-01190436637510.4103/wjnm.WJNM_4_20Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scannerElham Kashian0Hadi Ahangari1Vahab Dehlaghi2Karim Khoshgard3Pardis Ghafarian4Raheb Ghorbani5Department of Biomedical Engineering and Medical Physics, School of Medicine, Kermanshah University of Medical Sciences, KermanshahDepartment of Medical Physics, Faculty of medicine, Semnan University of Medical Sciences, SemnanDepartment of Biomedical Engineering and Medical Physics, School of Medicine, Kermanshah University of Medical Sciences, KermanshahDepartment of Medical Physics, School of Medicine, Kermanshah University of Medical Sciences, KermanshahChronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD)Department of Social Medicine, Faculty of medicine, Semnan University of Medical Sciences, SemnanThe aim of this study is to simulate GE Discovery 690 VCT positron emission tomography/computed tomography (PET/CT) scanner using Geant4 Application for Tomographic Emission (GATE) simulation package (version 8). Then, we assess the performance of scanner by comparing measured and simulated parameter results. Detection system and geometry of PET scanner that consists of 13,824 LYSO crystals designed in 256 blocks and 24 ring detectors were modeled. In order to achieve a precise model, we verified scanner model. Validation was based on a comparison between simulation data and experimental results obtained with this scanner in the same situation. Parameters used for validation were sensitivity, spatial resolution, and contrast. Image quality assessment was done based on comparing the contrast recovery coefficient (CRC) of simulated and measured images. The findings demonstrate that the mean difference between simulated and measured sensitivity is <7%. The simulated spatial resolution agreed to within <5.5% of the measured values. Contrast results had a slight divergence within the range below 4%. The image quality validation study demonstrated an acceptable agreement in CRC for 8:1 and 2:1 source-to-background activity ratio. Validated performance parameters showed good agreement between experimental data and simulated results and demonstrated that GATE is a valid simulation tool for simulating this scanner model. The simulated model of this scanner can be used for future studies regarding optimization of image reconstruction algorithms and emission acquisition protocols.http://www.thieme-connect.de/DOI/DOI?10.4103/wjnm.WJNM_4_20digital phantomgeant4 application for tomographic emissionmonte carlo simulationpositron emission tomography scannervalidation
spellingShingle Elham Kashian
Hadi Ahangari
Vahab Dehlaghi
Karim Khoshgard
Pardis Ghafarian
Raheb Ghorbani
Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner
World Journal of Nuclear Medicine
digital phantom
geant4 application for tomographic emission
monte carlo simulation
positron emission tomography scanner
validation
title Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner
title_full Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner
title_fullStr Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner
title_full_unstemmed Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner
title_short Monte Carlo simulation and performance assessment of GE Discovery 690 VCT positron emission tomography/computed tomography scanner
title_sort monte carlo simulation and performance assessment of ge discovery 690 vct positron emission tomography computed tomography scanner
topic digital phantom
geant4 application for tomographic emission
monte carlo simulation
positron emission tomography scanner
validation
url http://www.thieme-connect.de/DOI/DOI?10.4103/wjnm.WJNM_4_20
work_keys_str_mv AT elhamkashian montecarlosimulationandperformanceassessmentofgediscovery690vctpositronemissiontomographycomputedtomographyscanner
AT hadiahangari montecarlosimulationandperformanceassessmentofgediscovery690vctpositronemissiontomographycomputedtomographyscanner
AT vahabdehlaghi montecarlosimulationandperformanceassessmentofgediscovery690vctpositronemissiontomographycomputedtomographyscanner
AT karimkhoshgard montecarlosimulationandperformanceassessmentofgediscovery690vctpositronemissiontomographycomputedtomographyscanner
AT pardisghafarian montecarlosimulationandperformanceassessmentofgediscovery690vctpositronemissiontomographycomputedtomographyscanner
AT rahebghorbani montecarlosimulationandperformanceassessmentofgediscovery690vctpositronemissiontomographycomputedtomographyscanner