Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models.
<h4>Background</h4>This preclinical investigation evaluated the operational equivalence of a temporary intracoronary shunt (TICS) device while documenting patterns of endothelial repair during the perioperative period.<h4>Methods</h4>Porcine coronary bypass models were establ...
Saved in:
| Main Authors: | , , , , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Public Library of Science (PLoS)
2025-01-01
|
| Series: | PLoS ONE |
| Online Access: | https://doi.org/10.1371/journal.pone.0329864 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| _version_ | 1849228195674456064 |
|---|---|
| author | Qi Chen Yinfen Wang Liuhuan Huang Bin Li Zhuo Chen Ping Liu Xin Jiang |
| author_facet | Qi Chen Yinfen Wang Liuhuan Huang Bin Li Zhuo Chen Ping Liu Xin Jiang |
| author_sort | Qi Chen |
| collection | DOAJ |
| description | <h4>Background</h4>This preclinical investigation evaluated the operational equivalence of a temporary intracoronary shunt (TICS) device while documenting patterns of endothelial repair during the perioperative period.<h4>Methods</h4>Porcine coronary bypass models were established using off-pump (OPCAB) and on-pump beating-heart surgical approaches. Through a triple randomization design, test devices and predicate counterparts were deployed in left/right coronary arteries. Equivalence was determined at postoperative day 7 through tripartite analysis: 1) sequential laboratory measurements (preoperative to 7-day follow-up), 2) histopathological evaluation of vascular specimens, and 3) angiographic assessment. Extreme-sized device validation employed rabbit carotid arteries (small-caliber vascular bed) and abdominal aortae (large-caliber model), with equivalence criteria encompassing hemodynamic stability (flow velocity, trans-device pressure differentials) and microarchitectural preservation (endothelial integrity, internal elastic lamina continuity).<h4>Results</h4>In the conventional device cohort (12 target vessels), no perioperative type 5 myocardial infarction occurred. Postoperative CTA confirmed patent lumens and unobstructed distal flow in all vessels. Cardiac biomarkers (troponin, CK-MB, myoglobin) showed no significant differences at preoperative, 4h, 8h, 24h, 72h, or 7-day timepoints (P ≥ 0.05). For extreme-sized devices, hemodynamic parameters (mean proximal/distal pressure: test vs. predicate, P ≥ 0.05) and endothelial outcomes, including elastic lamina injury scores, demonstrated equivalence between test and predicate devices.<h4>Conclusion</h4>Functional parity between TICS and predicate devices was established in both porcine and lagomorph models, with observed endothelial alterations demonstrating transient characteristics limited to the acute perioperative window. |
| format | Article |
| id | doaj-art-f87f73b0df4e4740a2cf461a4aff0646 |
| institution | Kabale University |
| issn | 1932-6203 |
| language | English |
| publishDate | 2025-01-01 |
| publisher | Public Library of Science (PLoS) |
| record_format | Article |
| series | PLoS ONE |
| spelling | doaj-art-f87f73b0df4e4740a2cf461a4aff06462025-08-23T05:32:05ZengPublic Library of Science (PLoS)PLoS ONE1932-62032025-01-01208e032986410.1371/journal.pone.0329864Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models.Qi ChenYinfen WangLiuhuan HuangBin LiZhuo ChenPing LiuXin Jiang<h4>Background</h4>This preclinical investigation evaluated the operational equivalence of a temporary intracoronary shunt (TICS) device while documenting patterns of endothelial repair during the perioperative period.<h4>Methods</h4>Porcine coronary bypass models were established using off-pump (OPCAB) and on-pump beating-heart surgical approaches. Through a triple randomization design, test devices and predicate counterparts were deployed in left/right coronary arteries. Equivalence was determined at postoperative day 7 through tripartite analysis: 1) sequential laboratory measurements (preoperative to 7-day follow-up), 2) histopathological evaluation of vascular specimens, and 3) angiographic assessment. Extreme-sized device validation employed rabbit carotid arteries (small-caliber vascular bed) and abdominal aortae (large-caliber model), with equivalence criteria encompassing hemodynamic stability (flow velocity, trans-device pressure differentials) and microarchitectural preservation (endothelial integrity, internal elastic lamina continuity).<h4>Results</h4>In the conventional device cohort (12 target vessels), no perioperative type 5 myocardial infarction occurred. Postoperative CTA confirmed patent lumens and unobstructed distal flow in all vessels. Cardiac biomarkers (troponin, CK-MB, myoglobin) showed no significant differences at preoperative, 4h, 8h, 24h, 72h, or 7-day timepoints (P ≥ 0.05). For extreme-sized devices, hemodynamic parameters (mean proximal/distal pressure: test vs. predicate, P ≥ 0.05) and endothelial outcomes, including elastic lamina injury scores, demonstrated equivalence between test and predicate devices.<h4>Conclusion</h4>Functional parity between TICS and predicate devices was established in both porcine and lagomorph models, with observed endothelial alterations demonstrating transient characteristics limited to the acute perioperative window.https://doi.org/10.1371/journal.pone.0329864 |
| spellingShingle | Qi Chen Yinfen Wang Liuhuan Huang Bin Li Zhuo Chen Ping Liu Xin Jiang Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models. PLoS ONE |
| title | Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models. |
| title_full | Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models. |
| title_fullStr | Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models. |
| title_full_unstemmed | Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models. |
| title_short | Equivalence and endothelial safety of temporary intracoronary shunt device: Insights from preclinical porcine and rabbit models. |
| title_sort | equivalence and endothelial safety of temporary intracoronary shunt device insights from preclinical porcine and rabbit models |
| url | https://doi.org/10.1371/journal.pone.0329864 |
| work_keys_str_mv | AT qichen equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels AT yinfenwang equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels AT liuhuanhuang equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels AT binli equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels AT zhuochen equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels AT pingliu equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels AT xinjiang equivalenceandendothelialsafetyoftemporaryintracoronaryshuntdeviceinsightsfrompreclinicalporcineandrabbitmodels |