Ultrasonics absorption measurements of a liquid with backed transducer

An effect of backing of the transducer on the measured value of absorption of a liquid have been shown using shock excited system. The resulting deviation in the measured value of absorption from the accepted value has been observed to be large as 40 %, depending on the damping produced. A mechanism...

Full description

Saved in:
Bibliographic Details
Main Authors: V. N. BINDAL, T. K. SAKSENA, J. N. SOM
Format: Article
Language:English
Published: Institute of Fundamental Technological Research Polish Academy of Sciences 2015-07-01
Series:Archives of Acoustics
Online Access:https://acoustics.ippt.pan.pl/index.php/aa/article/view/3108
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An effect of backing of the transducer on the measured value of absorption of a liquid have been shown using shock excited system. The resulting deviation in the measured value of absorption from the accepted value has been observed to be large as 40 %, depending on the damping produced. A mechanism to explain the observed variation of the measured value of absorption with backing has been put forward. Implication of the work in characterisation of ultrasonic probe have been discussed.