Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard

Ying Zhang,1,* Ziyun Shi,2,* Haixia Liang,1 Meixia Chen,1 Yan Sun,1 Fujun Shang,3 Yi Wan4 1Department of Obstetrics and Gynecology, Xijing 986 hospital, Fourth Military Medical University, Xi’an, Shaanxi, People’s Republic of China; 2Department of Obstetrics, Shaanxi Provincial Peopl...

Full description

Saved in:
Bibliographic Details
Main Authors: Zhang Y, Shi Z, Liang H, Chen M, Sun Y, Shang F, Wan Y
Format: Article
Language:English
Published: Dove Medical Press 2024-11-01
Series:Vascular Health and Risk Management
Subjects:
Online Access:https://www.dovepress.com/clinical-accuracy-of-the-glab-md41a0-upper-arm-blood-pressure-monitor--peer-reviewed-fulltext-article-VHRM
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850179596185501696
author Zhang Y
Shi Z
Liang H
Chen M
Sun Y
Shang F
Wan Y
author_facet Zhang Y
Shi Z
Liang H
Chen M
Sun Y
Shang F
Wan Y
author_sort Zhang Y
collection DOAJ
description Ying Zhang,1,* Ziyun Shi,2,* Haixia Liang,1 Meixia Chen,1 Yan Sun,1 Fujun Shang,3 Yi Wan4 1Department of Obstetrics and Gynecology, Xijing 986 hospital, Fourth Military Medical University, Xi’an, Shaanxi, People’s Republic of China; 2Department of Obstetrics, Shaanxi Provincial People’s Hospital, Xi’an, Shaanxi, People’s Republic of China; 3Department of Cardiology, Xi’an Gem Flower Changqing Hospital, Xi’an, Shaanxi, People’s Republic of China; 4Department of Health Service, Fourth Military Medical University, Xi’an, Shaanxi, People’s Republic of China*These authors contributed equally to this workCorrespondence: Yi Wan, Department of Health Service, Fourth Military Medical University, No. 169 Changle Xilu, Xi’an, 710032, People’s Republic of China, Tel/Fax +86-29-84712676, Email wanyi@fmmu.edu.cn Fujun Shang, Department of Cardiology, Xi’an Gem Flower Changqing Hospital, Xi’an, Shaanxi, 710201, People’s Republic of China, Tel/Fax +86-29-86022756, Email shangfujun4@163.comObjective: Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060– 2:2018).Methods: A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.Results: The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was − 0.84± 3.88 mmHg and − 0.58± 3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.Conclusion: The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.Keywords: blood pressure monitor, validation, universal standard, pregnancy, pre-eclampsia
format Article
id doaj-art-ede967375c594b67b01ac1627f252cd8
institution OA Journals
issn 1178-2048
language English
publishDate 2024-11-01
publisher Dove Medical Press
record_format Article
series Vascular Health and Risk Management
spelling doaj-art-ede967375c594b67b01ac1627f252cd82025-08-20T02:18:27ZengDove Medical PressVascular Health and Risk Management1178-20482024-11-01Volume 2048749296946Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference StandardZhang YShi ZLiang HChen MSun YShang FWan YYing Zhang,1,* Ziyun Shi,2,* Haixia Liang,1 Meixia Chen,1 Yan Sun,1 Fujun Shang,3 Yi Wan4 1Department of Obstetrics and Gynecology, Xijing 986 hospital, Fourth Military Medical University, Xi’an, Shaanxi, People’s Republic of China; 2Department of Obstetrics, Shaanxi Provincial People’s Hospital, Xi’an, Shaanxi, People’s Republic of China; 3Department of Cardiology, Xi’an Gem Flower Changqing Hospital, Xi’an, Shaanxi, People’s Republic of China; 4Department of Health Service, Fourth Military Medical University, Xi’an, Shaanxi, People’s Republic of China*These authors contributed equally to this workCorrespondence: Yi Wan, Department of Health Service, Fourth Military Medical University, No. 169 Changle Xilu, Xi’an, 710032, People’s Republic of China, Tel/Fax +86-29-84712676, Email wanyi@fmmu.edu.cn Fujun Shang, Department of Cardiology, Xi’an Gem Flower Changqing Hospital, Xi’an, Shaanxi, 710201, People’s Republic of China, Tel/Fax +86-29-86022756, Email shangfujun4@163.comObjective: Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060– 2:2018).Methods: A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.Results: The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was − 0.84± 3.88 mmHg and − 0.58± 3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.Conclusion: The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.Keywords: blood pressure monitor, validation, universal standard, pregnancy, pre-eclampsiahttps://www.dovepress.com/clinical-accuracy-of-the-glab-md41a0-upper-arm-blood-pressure-monitor--peer-reviewed-fulltext-article-VHRMblood pressure monitorvalidationuniversal standardpregnancypre-eclampsia
spellingShingle Zhang Y
Shi Z
Liang H
Chen M
Sun Y
Shang F
Wan Y
Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard
Vascular Health and Risk Management
blood pressure monitor
validation
universal standard
pregnancy
pre-eclampsia
title Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard
title_full Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard
title_fullStr Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard
title_full_unstemmed Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard
title_short Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard
title_sort clinical accuracy of the g lab md41a0 upper arm blood pressure monitor in pregnancy and pre eclampsia using the aami esh iso as reference standard
topic blood pressure monitor
validation
universal standard
pregnancy
pre-eclampsia
url https://www.dovepress.com/clinical-accuracy-of-the-glab-md41a0-upper-arm-blood-pressure-monitor--peer-reviewed-fulltext-article-VHRM
work_keys_str_mv AT zhangy clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard
AT shiz clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard
AT liangh clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard
AT chenm clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard
AT suny clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard
AT shangf clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard
AT wany clinicalaccuracyoftheglabmd41a0upperarmbloodpressuremonitorinpregnancyandpreeclampsiausingtheaamieshisoasreferencestandard