Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
ABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardio...