Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases

ABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardio...

Full description

Saved in:
Bibliographic Details
Main Authors: Wael Zaher, Lorenzo Marcon, Antonio Sorgente
Format: Article
Language:English
Published: Wiley 2025-09-01
Series:Clinical Case Reports
Subjects:
Online Access:https://doi.org/10.1002/ccr3.70740
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849228047060828160
author Wael Zaher
Lorenzo Marcon
Antonio Sorgente
author_facet Wael Zaher
Lorenzo Marcon
Antonio Sorgente
author_sort Wael Zaher
collection DOAJ
description ABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardiovascular implantable electronic devices.
format Article
id doaj-art-e284faf9023b40f8a4d3ec674f956e5c
institution Kabale University
issn 2050-0904
language English
publishDate 2025-09-01
publisher Wiley
record_format Article
series Clinical Case Reports
spelling doaj-art-e284faf9023b40f8a4d3ec674f956e5c2025-08-23T07:09:50ZengWileyClinical Case Reports2050-09042025-09-01139n/an/a10.1002/ccr3.70740Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two CasesWael Zaher0Lorenzo Marcon1Antonio Sorgente2Department of Cardiology Centre Hospitalier EpiCURA Hornu Boussu BelgiumDepartment of Cardiology Centre Hospitalier EpiCURA Hornu Boussu BelgiumDepartment of Cardiology Centre Hospitalier EpiCURA Hornu Boussu BelgiumABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardiovascular implantable electronic devices.https://doi.org/10.1002/ccr3.70740cardiovascular implantable electronic deviceelectrical cardioversionelectromagnetic interferenceimplantable loop recorder
spellingShingle Wael Zaher
Lorenzo Marcon
Antonio Sorgente
Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
Clinical Case Reports
cardiovascular implantable electronic device
electrical cardioversion
electromagnetic interference
implantable loop recorder
title Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
title_full Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
title_fullStr Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
title_full_unstemmed Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
title_short Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
title_sort dysfunctional implantable loop recorder post electrical cardioversion a report of two cases
topic cardiovascular implantable electronic device
electrical cardioversion
electromagnetic interference
implantable loop recorder
url https://doi.org/10.1002/ccr3.70740
work_keys_str_mv AT waelzaher dysfunctionalimplantablelooprecorderpostelectricalcardioversionareportoftwocases
AT lorenzomarcon dysfunctionalimplantablelooprecorderpostelectricalcardioversionareportoftwocases
AT antoniosorgente dysfunctionalimplantablelooprecorderpostelectricalcardioversionareportoftwocases