Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases
ABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardio...
Saved in:
| Main Authors: | , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Wiley
2025-09-01
|
| Series: | Clinical Case Reports |
| Subjects: | |
| Online Access: | https://doi.org/10.1002/ccr3.70740 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| _version_ | 1849228047060828160 |
|---|---|
| author | Wael Zaher Lorenzo Marcon Antonio Sorgente |
| author_facet | Wael Zaher Lorenzo Marcon Antonio Sorgente |
| author_sort | Wael Zaher |
| collection | DOAJ |
| description | ABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardiovascular implantable electronic devices. |
| format | Article |
| id | doaj-art-e284faf9023b40f8a4d3ec674f956e5c |
| institution | Kabale University |
| issn | 2050-0904 |
| language | English |
| publishDate | 2025-09-01 |
| publisher | Wiley |
| record_format | Article |
| series | Clinical Case Reports |
| spelling | doaj-art-e284faf9023b40f8a4d3ec674f956e5c2025-08-23T07:09:50ZengWileyClinical Case Reports2050-09042025-09-01139n/an/a10.1002/ccr3.70740Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two CasesWael Zaher0Lorenzo Marcon1Antonio Sorgente2Department of Cardiology Centre Hospitalier EpiCURA Hornu Boussu BelgiumDepartment of Cardiology Centre Hospitalier EpiCURA Hornu Boussu BelgiumDepartment of Cardiology Centre Hospitalier EpiCURA Hornu Boussu BelgiumABSTRACT We report two occurrences of implantable loop recorder (ILR) dysfunction following electrical cardioversion, likely due to anterior–posterior patch positioning. Close proximity to the electrical field may affect ILRs, emphasizing the need for careful patch placement, similar to other cardiovascular implantable electronic devices.https://doi.org/10.1002/ccr3.70740cardiovascular implantable electronic deviceelectrical cardioversionelectromagnetic interferenceimplantable loop recorder |
| spellingShingle | Wael Zaher Lorenzo Marcon Antonio Sorgente Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases Clinical Case Reports cardiovascular implantable electronic device electrical cardioversion electromagnetic interference implantable loop recorder |
| title | Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases |
| title_full | Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases |
| title_fullStr | Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases |
| title_full_unstemmed | Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases |
| title_short | Dysfunctional Implantable Loop Recorder Post‐Electrical Cardioversion: A Report of Two Cases |
| title_sort | dysfunctional implantable loop recorder post electrical cardioversion a report of two cases |
| topic | cardiovascular implantable electronic device electrical cardioversion electromagnetic interference implantable loop recorder |
| url | https://doi.org/10.1002/ccr3.70740 |
| work_keys_str_mv | AT waelzaher dysfunctionalimplantablelooprecorderpostelectricalcardioversionareportoftwocases AT lorenzomarcon dysfunctionalimplantablelooprecorderpostelectricalcardioversionareportoftwocases AT antoniosorgente dysfunctionalimplantablelooprecorderpostelectricalcardioversionareportoftwocases |