Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures

Abstract Objective To explore the biomechanical properties of a four-dimensional structure locking osteosynthesis plate that can alter its structure and functionality over time. Methods According to the AO classification (AO-A2) standard, 18 artificial femoral shaft oblique fracture models with the...

Full description

Saved in:
Bibliographic Details
Main Authors: Meiling Song, Jian Wen, Zhe Wang, Yu Zeng, Jun Lu, Wei Lu, Changsheng Chen, Xieping Dong
Format: Article
Language:English
Published: BMC 2025-05-01
Series:BMC Musculoskeletal Disorders
Subjects:
Online Access:https://doi.org/10.1186/s12891-025-08756-z
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850105680677044224
author Meiling Song
Jian Wen
Zhe Wang
Yu Zeng
Jun Lu
Wei Lu
Changsheng Chen
Xieping Dong
author_facet Meiling Song
Jian Wen
Zhe Wang
Yu Zeng
Jun Lu
Wei Lu
Changsheng Chen
Xieping Dong
author_sort Meiling Song
collection DOAJ
description Abstract Objective To explore the biomechanical properties of a four-dimensional structure locking osteosynthesis plate that can alter its structure and functionality over time. Methods According to the AO classification (AO-A2) standard, 18 artificial femoral shaft oblique fracture models with the same position and shape were generated and subsequently randomly divided into 2 groups (n = 9). In the experimental group, the four-dimensional locking plate was fixed, and in the control group, the locking plate was fixed. In vitro loading experiments were carried out under three conditions: 1400 N axial compression, 500 N internal and external four-point bending, and 15 N·m torsion. The average compression stiffness, average bending stiffness, average torsional stiffness, and displacement were compared between the two groups. Results Under vertical loading, the average compression stiffness of the experimental group (558.7 ± 39.1) N/mm was greater than that of the control group (548.8 ± 24.5) N/mm, (P = 0.73); and the average displacement of the fracture end of the experimental group (3.3 ± 0.2) mm was lower than that of the control group (3.4 ± 0.3) mm, (P = 0.47). The differences were not statistically significant. Under four-point bending loading, the average bending stiffness of the experimental group (466.6 ± 85.8) N/mm was lower than that of the control group (542.5 ± 43.2) N/mm, and the difference was not statistically significant (P = 0.24). The average displacement of the fracture ends in the experimental group (1.6 ± 0.2)mm was greater than that in the control group (1.2 ± 0.1)mm, and the difference was statistically significant (P = 0.02). Under torsional loading, the average torsional stiffness of the experimental group (1.0 ± 0.1) N·m/deg was lower than that of the control group (1.1 ± 0.1) N·m/deg, (P = 0.27); and the average angular displacement of the fracture end of the experimental group (15.9 ± 1.3) deg was greater than that of the control group (14.6 ± 1.3) deg, (P = 0.26). There was no significant difference between the two groups. Conclusion The anti-compression, anti-bending and anti-torsion properties of four-dimensional locking plates are similar to those of conventional locking plates and can meet the mechanical requirements of internal fracture fixation.
format Article
id doaj-art-db74eadff882490e98daf46383c237eb
institution OA Journals
issn 1471-2474
language English
publishDate 2025-05-01
publisher BMC
record_format Article
series BMC Musculoskeletal Disorders
spelling doaj-art-db74eadff882490e98daf46383c237eb2025-08-20T02:39:01ZengBMCBMC Musculoskeletal Disorders1471-24742025-05-0126111110.1186/s12891-025-08756-zExperimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fracturesMeiling Song0Jian Wen1Zhe Wang2Yu Zeng3Jun Lu4Wei Lu5Changsheng Chen6Xieping Dong7 Jiangxi Provincial People’s Hospital, The First Affiliated Hospital of Nanchang Medical College, JXHC Key Laboratory of Digital Orthopedics Jiangxi Provincial People’s Hospital, The First Affiliated Hospital of Nanchang Medical College, JXHC Key Laboratory of Digital Orthopedics Jiangxi Provincial People’s Hospital, The First Affiliated Hospital of Nanchang Medical College, JXHC Key Laboratory of Digital Orthopedics Jiangxi Provincial People’s Hospital, The First Affiliated Hospital of Nanchang Medical College, JXHC Key Laboratory of Digital OrthopedicsSuzhou Kangli Orthopedics Instrument Co., Ltd.Suzhou Kangli Orthopedics Instrument Co., Ltd.Wuxi Times of Medical Science and Technology Co., Ltd. Jiangxi Provincial People’s Hospital, The First Affiliated Hospital of Nanchang Medical College, JXHC Key Laboratory of Digital OrthopedicsAbstract Objective To explore the biomechanical properties of a four-dimensional structure locking osteosynthesis plate that can alter its structure and functionality over time. Methods According to the AO classification (AO-A2) standard, 18 artificial femoral shaft oblique fracture models with the same position and shape were generated and subsequently randomly divided into 2 groups (n = 9). In the experimental group, the four-dimensional locking plate was fixed, and in the control group, the locking plate was fixed. In vitro loading experiments were carried out under three conditions: 1400 N axial compression, 500 N internal and external four-point bending, and 15 N·m torsion. The average compression stiffness, average bending stiffness, average torsional stiffness, and displacement were compared between the two groups. Results Under vertical loading, the average compression stiffness of the experimental group (558.7 ± 39.1) N/mm was greater than that of the control group (548.8 ± 24.5) N/mm, (P = 0.73); and the average displacement of the fracture end of the experimental group (3.3 ± 0.2) mm was lower than that of the control group (3.4 ± 0.3) mm, (P = 0.47). The differences were not statistically significant. Under four-point bending loading, the average bending stiffness of the experimental group (466.6 ± 85.8) N/mm was lower than that of the control group (542.5 ± 43.2) N/mm, and the difference was not statistically significant (P = 0.24). The average displacement of the fracture ends in the experimental group (1.6 ± 0.2)mm was greater than that in the control group (1.2 ± 0.1)mm, and the difference was statistically significant (P = 0.02). Under torsional loading, the average torsional stiffness of the experimental group (1.0 ± 0.1) N·m/deg was lower than that of the control group (1.1 ± 0.1) N·m/deg, (P = 0.27); and the average angular displacement of the fracture end of the experimental group (15.9 ± 1.3) deg was greater than that of the control group (14.6 ± 1.3) deg, (P = 0.26). There was no significant difference between the two groups. Conclusion The anti-compression, anti-bending and anti-torsion properties of four-dimensional locking plates are similar to those of conventional locking plates and can meet the mechanical requirements of internal fracture fixation.https://doi.org/10.1186/s12891-025-08756-zFemoral stem fractureFour-dimensional structure locking plateInternal fixationA-BO fixationBiomechanical loading
spellingShingle Meiling Song
Jian Wen
Zhe Wang
Yu Zeng
Jun Lu
Wei Lu
Changsheng Chen
Xieping Dong
Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures
BMC Musculoskeletal Disorders
Femoral stem fracture
Four-dimensional structure locking plate
Internal fixation
A-BO fixation
Biomechanical loading
title Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures
title_full Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures
title_fullStr Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures
title_full_unstemmed Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures
title_short Experimental study on the biomechanical loading of a four-dimensional plate for the fixation of femoral shaft fractures
title_sort experimental study on the biomechanical loading of a four dimensional plate for the fixation of femoral shaft fractures
topic Femoral stem fracture
Four-dimensional structure locking plate
Internal fixation
A-BO fixation
Biomechanical loading
url https://doi.org/10.1186/s12891-025-08756-z
work_keys_str_mv AT meilingsong experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT jianwen experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT zhewang experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT yuzeng experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT junlu experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT weilu experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT changshengchen experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures
AT xiepingdong experimentalstudyonthebiomechanicalloadingofafourdimensionalplateforthefixationoffemoralshaftfractures