Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration

Background. Traditional invasive hemoglobin (Hb) detection led to delayed diagnosis, operational inefficiency, incorrect critical decision making, and uncomfortable patient experience. To facilitate real-time total hemoglobin (tHb) monitoring, a portable prototype of a noninvasive Hb detection syste...

Full description

Saved in:
Bibliographic Details
Main Authors: Hongyun Liu, Fulai Peng, Minlu Hu, Jinlong Shi, Guojing Wang, Haiming Ai, Weidong Wang
Format: Article
Language:English
Published: Wiley 2020-01-01
Series:Journal of Electrical and Computer Engineering
Online Access:http://dx.doi.org/10.1155/2020/3034260
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832546907998650368
author Hongyun Liu
Fulai Peng
Minlu Hu
Jinlong Shi
Guojing Wang
Haiming Ai
Weidong Wang
author_facet Hongyun Liu
Fulai Peng
Minlu Hu
Jinlong Shi
Guojing Wang
Haiming Ai
Weidong Wang
author_sort Hongyun Liu
collection DOAJ
description Background. Traditional invasive hemoglobin (Hb) detection led to delayed diagnosis, operational inefficiency, incorrect critical decision making, and uncomfortable patient experience. To facilitate real-time total hemoglobin (tHb) monitoring, a portable prototype of a noninvasive Hb detection system was developed, and the accuracy of Hb predicted based on partial least squares (PLS) as well as backpropagation artificial neural network (BP-ANN) models was validated. Results. The prototype was combined with a signal processing circuit and a spectrophotometric probe containing 8 wavelength LEDs light source and photodiode array. Laboratory invasive Hb (Lab_tHb) and spot check Hb measurements with PLS (SpHb_PLS) and BP (SpHb_BP) methods were obtained simultaneously by hematology analyzer and the designed system. The invasive and noninvasive estimates of the Hb levels were analyzed using Spearman correlation as well as Bland–Altman plot and receiver operating characteristic (ROC) curve analysis. A total of 238 volunteers had attempted laboratory invasive and noninvasive spot check Hb measurements. Mean Lab_tHb, SpHb_PLS, and SpHb_BP were 13.6 ± 1.80 g/dL, 13.5 ± 1.07 g/dL, and 13.6 ± 1.06 g/dL, respectively. Noninvasive SpHb_PLS (r = 0.61, p<0.001) and SpHb_BP (r = 0.62, p<0.001) had a strong correlation with invasive tHb values. The Bland–Altman plot showed excellent consistency between the proposed noninvasive methods and laboratory invasive reference. In ROC analysis, PLS and BP models were good at predicting Hb ≥ 12 g/dL with area under the curve of 0.828 and 0.824, respectively. Observed differences between invasive and noninvasive Hb measurements displayed no significant correlation with perfusion index values. Conclusions. The result confirmed that noninvasive Hb monitoring had an excellent correlation with traditional invasive Hb measurement. Furthermore, it is suggested that the developed prototype has the potential for the noninvasive detection of Hb concentration with the methods of PLS and BP-ANN.
format Article
id doaj-art-d93626294c6e4348ae5c4875a0e7a7de
institution Kabale University
issn 2090-0147
2090-0155
language English
publishDate 2020-01-01
publisher Wiley
record_format Article
series Journal of Electrical and Computer Engineering
spelling doaj-art-d93626294c6e4348ae5c4875a0e7a7de2025-02-03T06:46:53ZengWileyJournal of Electrical and Computer Engineering2090-01472090-01552020-01-01202010.1155/2020/30342603034260Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin ConcentrationHongyun Liu0Fulai Peng1Minlu Hu2Jinlong Shi3Guojing Wang4Haiming Ai5Weidong Wang6Department of Biomedical Engineering, Chinese PLA General Hospital, Beijing 100853, ChinaShandong Institute of Advanced Technology, Chinese Academy of Sciences, Beijing 250000, ChinaCenter of Medical Device R & D and Clinical Evaluation, Chinese PLA General Hospital, Beijing 100853, ChinaDepartment of Biomedical Engineering, Chinese PLA General Hospital, Beijing 100853, ChinaDepartment of Biomedical Engineering, Chinese PLA General Hospital, Beijing 100853, ChinaDepartment of Computer Science, Beijing Open University, Beijing 100081, ChinaDepartment of Biomedical Engineering, Chinese PLA General Hospital, Beijing 100853, ChinaBackground. Traditional invasive hemoglobin (Hb) detection led to delayed diagnosis, operational inefficiency, incorrect critical decision making, and uncomfortable patient experience. To facilitate real-time total hemoglobin (tHb) monitoring, a portable prototype of a noninvasive Hb detection system was developed, and the accuracy of Hb predicted based on partial least squares (PLS) as well as backpropagation artificial neural network (BP-ANN) models was validated. Results. The prototype was combined with a signal processing circuit and a spectrophotometric probe containing 8 wavelength LEDs light source and photodiode array. Laboratory invasive Hb (Lab_tHb) and spot check Hb measurements with PLS (SpHb_PLS) and BP (SpHb_BP) methods were obtained simultaneously by hematology analyzer and the designed system. The invasive and noninvasive estimates of the Hb levels were analyzed using Spearman correlation as well as Bland–Altman plot and receiver operating characteristic (ROC) curve analysis. A total of 238 volunteers had attempted laboratory invasive and noninvasive spot check Hb measurements. Mean Lab_tHb, SpHb_PLS, and SpHb_BP were 13.6 ± 1.80 g/dL, 13.5 ± 1.07 g/dL, and 13.6 ± 1.06 g/dL, respectively. Noninvasive SpHb_PLS (r = 0.61, p<0.001) and SpHb_BP (r = 0.62, p<0.001) had a strong correlation with invasive tHb values. The Bland–Altman plot showed excellent consistency between the proposed noninvasive methods and laboratory invasive reference. In ROC analysis, PLS and BP models were good at predicting Hb ≥ 12 g/dL with area under the curve of 0.828 and 0.824, respectively. Observed differences between invasive and noninvasive Hb measurements displayed no significant correlation with perfusion index values. Conclusions. The result confirmed that noninvasive Hb monitoring had an excellent correlation with traditional invasive Hb measurement. Furthermore, it is suggested that the developed prototype has the potential for the noninvasive detection of Hb concentration with the methods of PLS and BP-ANN.http://dx.doi.org/10.1155/2020/3034260
spellingShingle Hongyun Liu
Fulai Peng
Minlu Hu
Jinlong Shi
Guojing Wang
Haiming Ai
Weidong Wang
Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
Journal of Electrical and Computer Engineering
title Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
title_full Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
title_fullStr Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
title_full_unstemmed Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
title_short Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
title_sort development and validation of a photoplethysmography system for noninvasive monitoring of hemoglobin concentration
url http://dx.doi.org/10.1155/2020/3034260
work_keys_str_mv AT hongyunliu developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration
AT fulaipeng developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration
AT minluhu developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration
AT jinlongshi developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration
AT guojingwang developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration
AT haimingai developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration
AT weidongwang developmentandvalidationofaphotoplethysmographysystemfornoninvasivemonitoringofhemoglobinconcentration