Iris-Occlusion of XEN Gel Stent following Ab Externo Transconjunctival Implantation Technique

The use of minimally invasive glaucoma surgery (MIGS) devices has become increasingly common for the management of elevated intraocular pressure (IOP) in the context of glaucoma. These technologies have traditionally been associated with fewer postoperative complications than conventional surgical t...

Full description

Saved in:
Bibliographic Details
Main Authors: Sunil Ruparelia, Nir Shoham-Hazon
Format: Article
Language:English
Published: Wiley 2021-01-01
Series:Case Reports in Ophthalmological Medicine
Online Access:http://dx.doi.org/10.1155/2021/2936047
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The use of minimally invasive glaucoma surgery (MIGS) devices has become increasingly common for the management of elevated intraocular pressure (IOP) in the context of glaucoma. These technologies have traditionally been associated with fewer postoperative complications than conventional surgical techniques. However, we report on a rare case of transient XEN occlusion associated with pupil dilation following XEN gel stent implantation. This case highlights that in future XEN implantations, it may be preferable to position the XEN at a lesser angle to the iris to prevent such an occlusion. The use of different positionings of XEN is performed to optimize outcomes. However, it is highlighted that complications may arise in certain circumstances.
ISSN:2090-6722