The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)

The aim of the study. To study the effect of high nitric oxide concentrations on hollow polypropylene fibers of oxygenators.Materials and methods. The study was conducted in two stages. At the first stage, we evaluated the stability of oxygenator membrane made of hollow polypropylene fibers after si...

Full description

Saved in:
Bibliographic Details
Main Authors: A. M. Radovskiy, I. V. Vorotyntsev, A. A. Atlaskin, A. N. Petukhov, S. S. Kryuchkov, M. E. Atlaskina, A. N. Stepakova, A. O. Marichev, E. K. Barygin, V. V. Osovskikh, V. D. Selemir, S. N. Buranov, V. V. Golovanov, A. S. Shirshin, Yu. V. Valueva, V. V. Pichugin, S. E. Domnin, A. E. Bautin
Format: Article
Language:English
Published: Federal Research and Clinical Center of Intensive Care Medicine and Rehabilitology, Moscow, Russia 2024-02-01
Series:Общая реаниматология
Subjects:
Online Access:https://www.reanimatology.com/rmt/article/view/2351
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849242789618909184
author A. M. Radovskiy
I. V. Vorotyntsev
A. A. Atlaskin
A. N. Petukhov
S. S. Kryuchkov
M. E. Atlaskina
A. N. Stepakova
A. O. Marichev
E. K. Barygin
V. V. Osovskikh
V. D. Selemir
S. N. Buranov
V. V. Golovanov
A. S. Shirshin
Yu. V. Valueva
V. V. Pichugin
S. E. Domnin
A. E. Bautin
author_facet A. M. Radovskiy
I. V. Vorotyntsev
A. A. Atlaskin
A. N. Petukhov
S. S. Kryuchkov
M. E. Atlaskina
A. N. Stepakova
A. O. Marichev
E. K. Barygin
V. V. Osovskikh
V. D. Selemir
S. N. Buranov
V. V. Golovanov
A. S. Shirshin
Yu. V. Valueva
V. V. Pichugin
S. E. Domnin
A. E. Bautin
author_sort A. M. Radovskiy
collection DOAJ
description The aim of the study. To study the effect of high nitric oxide concentrations on hollow polypropylene fibers of oxygenators.Materials and methods. The study was conducted in two stages. At the first stage, we evaluated the stability of oxygenator membrane made of hollow polypropylene fibers after six hours of exposure to air-oxygen mixture containing NO at 500 parts per million, or 500 pro pro mille (ppm) concentration, using mass spectrometry and infrared spectroscopy. At the second stage, an experiment with cardiopulmonary bypass (CPB) was conducted on 10 pigs. In the study group (n=5) animals sweep gas was supplied to the oxygenator as an air-oxygen mixture with NO at 100 ppm. In the control group animals (n=5) an air-oxygen mixture was used without NO. The CPB lasted for 4 hours, followed by observation for 12 hours. NO, NO2 (at the inlet and outlet of the oxygenator), and the dynamics of methemoglobin were evaluated. After weaning of animals from CPB, the oxygenators were tested for leakproofness, and scanning electron microscopy (SEM) was performed.Results. The oxygenator made of polypropylene hollow fibers retained its gas transfer parameters after six hours of exposure to air-oxygen mixture containing NO at 500 ppm. Based on IR-Fourier spectroscopy findings, NO did not affect structural integrity of polypropylene membranes. NO added to gas mixture at 100 ppm did not increase NO2 to toxic level of 2 ppm in 91% of control tests during 4 hours CPB in pigs; mean value was 1.58 ± 0.28 ppm. Methemoglobin concentration did not exceed the upper limit of permissible level (3%), and there were no statistically significant differences with the control group. All tested oxygenators have passed the leakproofness test. According to SEM findings, larger amounts of fibrin deposits were found in the control group oxygenators vs study group.Conclusion. There were no negative effects of NO at 500 ppm concentration on the oxygenator membrane made of hollow polypropylene fibers. NO at 100 ppm in a gas-mixture supplied to oxygenators did not lead to an exceedance of safe NO2 and methemoglobin concentrations in an animal model. Reduced fibrin deposits on hollow fibers of polypropylene oxygenator membranes were observed when with NO at a level of 100 ppm was added to a gas mixture.
format Article
id doaj-art-d3cb7dea74a34924913d2355b3d7787a
institution Kabale University
issn 1813-9779
2411-7110
language English
publishDate 2024-02-01
publisher Federal Research and Clinical Center of Intensive Care Medicine and Rehabilitology, Moscow, Russia
record_format Article
series Общая реаниматология
spelling doaj-art-d3cb7dea74a34924913d2355b3d7787a2025-08-20T03:59:40ZengFederal Research and Clinical Center of Intensive Care Medicine and Rehabilitology, Moscow, RussiaОбщая реаниматология1813-97792411-71102024-02-01201506210.15360/1813-9779-2024-1-23511928The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)A. M. Radovskiy0I. V. Vorotyntsev1A. A. Atlaskin2A. N. Petukhov3S. S. Kryuchkov4M. E. Atlaskina5A. N. Stepakova6A. O. Marichev7E. K. Barygin8V. V. Osovskikh9V. D. Selemir10S. N. Buranov11V. V. Golovanov12A. S. Shirshin13Yu. V. Valueva14V. V. Pichugin15S. E. Domnin16A. E. Bautin17Almazov National Medical Research CenterMendeleev University of Chemical Technology of RussiaMendeleev University of Chemical Technology of RussiaMendeleev University of Chemical Technology of RussiaMendeleev University of Chemical Technology of RussiaMendeleev University of Chemical Technology of RussiaMendeleev University of Chemical Technology of RussiaAlmazov National Medical Research CenterAlmazov National Medical Research CenterAlmazov National Medical Research CenterRussian Federal Nuclear Center, All-Russian Scientific Research Institute of Experimental PhysicsRussian Federal Nuclear Center, All-Russian Scientific Research Institute of Experimental PhysicsRussian Federal Nuclear Center, All-Russian Scientific Research Institute of Experimental PhysicsRussian Federal Nuclear Center, All-Russian Scientific Research Institute of Experimental PhysicsRussian Federal Nuclear Center, All-Russian Scientific Research Institute of Experimental PhysicsPrivolzhsky Research Medical UniversityResearch Institute “Specialized Cardiosurgical Clinical Hospital named after Academician B.A. Koroleva”Almazov National Medical Research CenterThe aim of the study. To study the effect of high nitric oxide concentrations on hollow polypropylene fibers of oxygenators.Materials and methods. The study was conducted in two stages. At the first stage, we evaluated the stability of oxygenator membrane made of hollow polypropylene fibers after six hours of exposure to air-oxygen mixture containing NO at 500 parts per million, or 500 pro pro mille (ppm) concentration, using mass spectrometry and infrared spectroscopy. At the second stage, an experiment with cardiopulmonary bypass (CPB) was conducted on 10 pigs. In the study group (n=5) animals sweep gas was supplied to the oxygenator as an air-oxygen mixture with NO at 100 ppm. In the control group animals (n=5) an air-oxygen mixture was used without NO. The CPB lasted for 4 hours, followed by observation for 12 hours. NO, NO2 (at the inlet and outlet of the oxygenator), and the dynamics of methemoglobin were evaluated. After weaning of animals from CPB, the oxygenators were tested for leakproofness, and scanning electron microscopy (SEM) was performed.Results. The oxygenator made of polypropylene hollow fibers retained its gas transfer parameters after six hours of exposure to air-oxygen mixture containing NO at 500 ppm. Based on IR-Fourier spectroscopy findings, NO did not affect structural integrity of polypropylene membranes. NO added to gas mixture at 100 ppm did not increase NO2 to toxic level of 2 ppm in 91% of control tests during 4 hours CPB in pigs; mean value was 1.58 ± 0.28 ppm. Methemoglobin concentration did not exceed the upper limit of permissible level (3%), and there were no statistically significant differences with the control group. All tested oxygenators have passed the leakproofness test. According to SEM findings, larger amounts of fibrin deposits were found in the control group oxygenators vs study group.Conclusion. There were no negative effects of NO at 500 ppm concentration on the oxygenator membrane made of hollow polypropylene fibers. NO at 100 ppm in a gas-mixture supplied to oxygenators did not lead to an exceedance of safe NO2 and methemoglobin concentrations in an animal model. Reduced fibrin deposits on hollow fibers of polypropylene oxygenator membranes were observed when with NO at a level of 100 ppm was added to a gas mixture.https://www.reanimatology.com/rmt/article/view/2351nitric oxidecardio-pulmonary bypassoxygenatorpolypropylene hollow fiberscardiac surgery
spellingShingle A. M. Radovskiy
I. V. Vorotyntsev
A. A. Atlaskin
A. N. Petukhov
S. S. Kryuchkov
M. E. Atlaskina
A. N. Stepakova
A. O. Marichev
E. K. Barygin
V. V. Osovskikh
V. D. Selemir
S. N. Buranov
V. V. Golovanov
A. S. Shirshin
Yu. V. Valueva
V. V. Pichugin
S. E. Domnin
A. E. Bautin
The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)
Общая реаниматология
nitric oxide
cardio-pulmonary bypass
oxygenator
polypropylene hollow fibers
cardiac surgery
title The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)
title_full The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)
title_fullStr The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)
title_full_unstemmed The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)
title_short The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)
title_sort effect of high nitric oxide concentrations on oxygenators in cardiopulmonary bypass machines experimental study
topic nitric oxide
cardio-pulmonary bypass
oxygenator
polypropylene hollow fibers
cardiac surgery
url https://www.reanimatology.com/rmt/article/view/2351
work_keys_str_mv AT amradovskiy theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT ivvorotyntsev theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT aaatlaskin theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT anpetukhov theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT sskryuchkov theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT meatlaskina theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT anstepakova theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT aomarichev theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT ekbarygin theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vvosovskikh theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vdselemir theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT snburanov theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vvgolovanov theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT asshirshin theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT yuvvalueva theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vvpichugin theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT sedomnin theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT aebautin theeffectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT amradovskiy effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT ivvorotyntsev effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT aaatlaskin effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT anpetukhov effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT sskryuchkov effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT meatlaskina effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT anstepakova effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT aomarichev effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT ekbarygin effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vvosovskikh effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vdselemir effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT snburanov effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vvgolovanov effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT asshirshin effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT yuvvalueva effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT vvpichugin effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT sedomnin effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy
AT aebautin effectofhighnitricoxideconcentrationsonoxygenatorsincardiopulmonarybypassmachinesexperimentalstudy