Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom

Purpose. We sought to examine the possibility of reducing the contrast medium dosage in dual-energy imaging using a saline-mixed injection with a virtual monochromatic energy method of dual-source computed tomography (CT). Methods. An X-ray CT (SOMATOM Definition Flash: Siemens, Nurnberg, Germany) w...

Full description

Saved in:
Bibliographic Details
Main Authors: Hironobu Tomita, Koichi Shibata
Format: Article
Language:English
Published: Wiley 2019-01-01
Series:Radiology Research and Practice
Online Access:http://dx.doi.org/10.1155/2019/3512126
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850175304072429568
author Hironobu Tomita
Koichi Shibata
author_facet Hironobu Tomita
Koichi Shibata
author_sort Hironobu Tomita
collection DOAJ
description Purpose. We sought to examine the possibility of reducing the contrast medium dosage in dual-energy imaging using a saline-mixed injection with a virtual monochromatic energy method of dual-source computed tomography (CT). Methods. An X-ray CT (SOMATOM Definition Flash: Siemens, Nurnberg, Germany) was employed. The mixing ratio of contrast medium and saline was gradually changed by 10%, followed by a mixed injection into a dynamic blood flow phantom (Nemoto Kyorindo, Japan) which is a hemodynamic simulation phantom to obtain time-enhancement curves (TECs). Exactly 64 TECs were prepared for each mixing ratio by changing the energy from 40 to 75 keV for monoenergetic imaging. The relationship between the image standard deviation (SD) and the energy of the virtual monochromatic image was determined. Combinations of the mixing ratio and energy (keV), which can maintain high CT numbers and low image SDs for 3D imaging, were tested, and the reduction rate of the contrast medium was calculated. Results. The TECs for the mixed injection method changed linearly with the dilution rates. The mixing ratios were strongly correlated with the maximum CT number of the TEC (R2 = 0.98). Contrast CT numbers and image SDs increased by approximately 20% and 25%, respectively, as the energy decreased by 5 keV. The optimal conditions for reducing the contrast medium dose were a mixing ratio of 6:4 and 55 keV of energy. Conclusion. The virtual monochromatic energy method reduced the contrast medium dosage by up to 40% for three-dimensional CT-angio (3DCTA) tests.
format Article
id doaj-art-cef0451802c44be59a5463ebaaf84ebd
institution OA Journals
issn 2090-1941
2090-195X
language English
publishDate 2019-01-01
publisher Wiley
record_format Article
series Radiology Research and Practice
spelling doaj-art-cef0451802c44be59a5463ebaaf84ebd2025-08-20T02:19:30ZengWileyRadiology Research and Practice2090-19412090-195X2019-01-01201910.1155/2019/35121263512126Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow PhantomHironobu Tomita0Koichi Shibata1Graduate School of Healthcare Science, Suzuka University of Medical Science, 1001-1, Kishioka-cho, Suzuka, Mie 510-0293, JapanGraduate School of Healthcare Science, Suzuka University of Medical Science, 1001-1, Kishioka-cho, Suzuka, Mie 510-0293, JapanPurpose. We sought to examine the possibility of reducing the contrast medium dosage in dual-energy imaging using a saline-mixed injection with a virtual monochromatic energy method of dual-source computed tomography (CT). Methods. An X-ray CT (SOMATOM Definition Flash: Siemens, Nurnberg, Germany) was employed. The mixing ratio of contrast medium and saline was gradually changed by 10%, followed by a mixed injection into a dynamic blood flow phantom (Nemoto Kyorindo, Japan) which is a hemodynamic simulation phantom to obtain time-enhancement curves (TECs). Exactly 64 TECs were prepared for each mixing ratio by changing the energy from 40 to 75 keV for monoenergetic imaging. The relationship between the image standard deviation (SD) and the energy of the virtual monochromatic image was determined. Combinations of the mixing ratio and energy (keV), which can maintain high CT numbers and low image SDs for 3D imaging, were tested, and the reduction rate of the contrast medium was calculated. Results. The TECs for the mixed injection method changed linearly with the dilution rates. The mixing ratios were strongly correlated with the maximum CT number of the TEC (R2 = 0.98). Contrast CT numbers and image SDs increased by approximately 20% and 25%, respectively, as the energy decreased by 5 keV. The optimal conditions for reducing the contrast medium dose were a mixing ratio of 6:4 and 55 keV of energy. Conclusion. The virtual monochromatic energy method reduced the contrast medium dosage by up to 40% for three-dimensional CT-angio (3DCTA) tests.http://dx.doi.org/10.1155/2019/3512126
spellingShingle Hironobu Tomita
Koichi Shibata
Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom
Radiology Research and Practice
title Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom
title_full Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom
title_fullStr Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom
title_full_unstemmed Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom
title_short Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom
title_sort verification of the dose reduction effect via diluted injection in dual energy computed tomography using a human blood flow phantom
url http://dx.doi.org/10.1155/2019/3512126
work_keys_str_mv AT hironobutomita verificationofthedosereductioneffectviadilutedinjectionindualenergycomputedtomographyusingahumanbloodflowphantom
AT koichishibata verificationofthedosereductioneffectviadilutedinjectionindualenergycomputedtomographyusingahumanbloodflowphantom