Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults

Herkko Ryynänen,1,2 Anssi Sovijärvi,3 Ilpo Kuronen,4 Essi K Ahokas,1 Maarit Valtonen,5 Johanna K Ihalainen,1 Juhani Multanen1,6 1Faculty of Sport and Health Sciences, University of Jyväskylä, Jyväskylä, Finland; 2Finnish Lung Health Association, Helsinki, Finland; 3Department of Clinical Physiology,...

Full description

Saved in:
Bibliographic Details
Main Authors: Ryynänen H, Sovijärvi A, Kuronen I, Ahokas EK, Valtonen M, Ihalainen JK, Multanen J
Format: Article
Language:English
Published: Dove Medical Press 2025-06-01
Series:Medical Devices: Evidence and Research
Subjects:
Online Access:https://www.dovepress.com/validity-and-short-term-repeatability-of-a-novel-hand-held-respiratory-peer-reviewed-fulltext-article-MDER
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850157386100113408
author Ryynänen H
Sovijärvi A
Kuronen I
Ahokas EK
Valtonen M
Ihalainen JK
Multanen J
author_facet Ryynänen H
Sovijärvi A
Kuronen I
Ahokas EK
Valtonen M
Ihalainen JK
Multanen J
author_sort Ryynänen H
collection DOAJ
description Herkko Ryynänen,1,2 Anssi Sovijärvi,3 Ilpo Kuronen,4 Essi K Ahokas,1 Maarit Valtonen,5 Johanna K Ihalainen,1 Juhani Multanen1,6 1Faculty of Sport and Health Sciences, University of Jyväskylä, Jyväskylä, Finland; 2Finnish Lung Health Association, Helsinki, Finland; 3Department of Clinical Physiology, University of Helsinki, Helsinki, Finland; 4Reagena Ltd., Toivala, Finland; 5Finnish Institute of High Performance Sport KIHU, Jyväskylä, Finland; 6South-Eastern Finland University of Applied Sciences, Savonlinna, FinlandCorrespondence: Herkko Ryynänen, Faculty of Sport and Health Sciences, University of Jyväskylä, PO Box 35, Jyväskylä, Finland, Tel +358 40 7030231, Email herkko.ryynanen@gmail.comPurpose: Measuring maximal airway pressure is an essential part of the assessment of respiratory functions. Portable handheld devices have made clinical measurements more available, but reliable and user-friendly devices for non-clinical use remain rare. This study sought to determine the validity and short-term repeatability of measurements of dynamic maximal inspiratory pressure (dMIP) and dynamic maximal expiratory pressure (dMEP) by using a novel self-administered respiratory health meter (WellO2-RHM) in asymptomatic young adults.Patients and Methods: dMIP and dMEP were measured with WellO2-RHM in asymptomatic adult volunteers (n=26, 15 male and 11 female, age 26– 41 years). These values were compared with quasi-static maximal inspiratory pressure (MIP) and maximal expiratory pressure (MEP) obtained from the same volunteers using another respiratory manometer (MicroRPM). The measurements of dMIP and dMEP with WellO2-RHM were repeated in the same individuals at an interval of one week for assessment of their short-term repeatability.Results: The Pearson correlation coefficients of dMIP and dMEP values with MIP and MEP values were high (r=0.840, p< 0.001; r=0.849, p< 0.001, respectively). The dMIP and dMEP values were consistently lower than the quasi-static MIP and MEP values in the same individuals. The short-term repeatability of the dMIP and dMEP in one week interval proved to be moderately good in terms of the coefficient of variation (CV), the intraclass correlation coefficient (ICC), the standard error of measurement (SEm) and minimal detectable change (MDC) (10.0%, 0.825, p< 0.001, 7 cmH2O and 20 cmH2O and 9.1%, 0.895, p< 0.001, 12 cmH2O and 34 cmH2O, respectively).Conclusion: The results indicate that the WellO2-RHM is a valid and repeatable method for the assessment of dMIP and dMEP in asymptomatic young adults, but the absolute values are lower than those obtained with devices measuring quasi-static MIP and MEP. The findings suggest that WellO2-RHM can be used for self-monitoring of the effects of respiratory muscle training in healthy young adults.Keywords: breathing, respiratory muscles, assessment, maximal airway pressures, short-term repeatability
format Article
id doaj-art-ce73e6fe2ecc40f89230d23243b5a419
institution OA Journals
issn 1179-1470
language English
publishDate 2025-06-01
publisher Dove Medical Press
record_format Article
series Medical Devices: Evidence and Research
spelling doaj-art-ce73e6fe2ecc40f89230d23243b5a4192025-08-20T02:24:13ZengDove Medical PressMedical Devices: Evidence and Research1179-14702025-06-01Volume 18Issue 1297308103633Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young AdultsRyynänen HSovijärvi A0Kuronen I1Ahokas EK2Valtonen MIhalainen JK3Multanen J4Clinical PhysiologyR&DFaculty of Sport and Health SciencesFaculty of Sport and Health SciencesDepartment of Social Services, Health and SportsHerkko Ryynänen,1,2 Anssi Sovijärvi,3 Ilpo Kuronen,4 Essi K Ahokas,1 Maarit Valtonen,5 Johanna K Ihalainen,1 Juhani Multanen1,6 1Faculty of Sport and Health Sciences, University of Jyväskylä, Jyväskylä, Finland; 2Finnish Lung Health Association, Helsinki, Finland; 3Department of Clinical Physiology, University of Helsinki, Helsinki, Finland; 4Reagena Ltd., Toivala, Finland; 5Finnish Institute of High Performance Sport KIHU, Jyväskylä, Finland; 6South-Eastern Finland University of Applied Sciences, Savonlinna, FinlandCorrespondence: Herkko Ryynänen, Faculty of Sport and Health Sciences, University of Jyväskylä, PO Box 35, Jyväskylä, Finland, Tel +358 40 7030231, Email herkko.ryynanen@gmail.comPurpose: Measuring maximal airway pressure is an essential part of the assessment of respiratory functions. Portable handheld devices have made clinical measurements more available, but reliable and user-friendly devices for non-clinical use remain rare. This study sought to determine the validity and short-term repeatability of measurements of dynamic maximal inspiratory pressure (dMIP) and dynamic maximal expiratory pressure (dMEP) by using a novel self-administered respiratory health meter (WellO2-RHM) in asymptomatic young adults.Patients and Methods: dMIP and dMEP were measured with WellO2-RHM in asymptomatic adult volunteers (n=26, 15 male and 11 female, age 26– 41 years). These values were compared with quasi-static maximal inspiratory pressure (MIP) and maximal expiratory pressure (MEP) obtained from the same volunteers using another respiratory manometer (MicroRPM). The measurements of dMIP and dMEP with WellO2-RHM were repeated in the same individuals at an interval of one week for assessment of their short-term repeatability.Results: The Pearson correlation coefficients of dMIP and dMEP values with MIP and MEP values were high (r=0.840, p< 0.001; r=0.849, p< 0.001, respectively). The dMIP and dMEP values were consistently lower than the quasi-static MIP and MEP values in the same individuals. The short-term repeatability of the dMIP and dMEP in one week interval proved to be moderately good in terms of the coefficient of variation (CV), the intraclass correlation coefficient (ICC), the standard error of measurement (SEm) and minimal detectable change (MDC) (10.0%, 0.825, p< 0.001, 7 cmH2O and 20 cmH2O and 9.1%, 0.895, p< 0.001, 12 cmH2O and 34 cmH2O, respectively).Conclusion: The results indicate that the WellO2-RHM is a valid and repeatable method for the assessment of dMIP and dMEP in asymptomatic young adults, but the absolute values are lower than those obtained with devices measuring quasi-static MIP and MEP. The findings suggest that WellO2-RHM can be used for self-monitoring of the effects of respiratory muscle training in healthy young adults.Keywords: breathing, respiratory muscles, assessment, maximal airway pressures, short-term repeatabilityhttps://www.dovepress.com/validity-and-short-term-repeatability-of-a-novel-hand-held-respiratory-peer-reviewed-fulltext-article-MDERbreathingrespiratory musclesassessmentmaximal airway pressuresshort-term repeatability.
spellingShingle Ryynänen H
Sovijärvi A
Kuronen I
Ahokas EK
Valtonen M
Ihalainen JK
Multanen J
Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults
Medical Devices: Evidence and Research
breathing
respiratory muscles
assessment
maximal airway pressures
short-term repeatability.
title Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults
title_full Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults
title_fullStr Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults
title_full_unstemmed Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults
title_short Validity and Short-Term Repeatability of a Novel Hand-Held Respiratory Health Meter for the Assessment of Dynamic Maximal Respiratory Pressures in Healthy Young Adults
title_sort validity and short term repeatability of a novel hand held respiratory health meter for the assessment of dynamic maximal respiratory pressures in healthy young adults
topic breathing
respiratory muscles
assessment
maximal airway pressures
short-term repeatability.
url https://www.dovepress.com/validity-and-short-term-repeatability-of-a-novel-hand-held-respiratory-peer-reviewed-fulltext-article-MDER
work_keys_str_mv AT ryynanenh validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults
AT sovijarvia validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults
AT kuroneni validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults
AT ahokasek validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults
AT valtonenm validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults
AT ihalainenjk validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults
AT multanenj validityandshorttermrepeatabilityofanovelhandheldrespiratoryhealthmeterfortheassessmentofdynamicmaximalrespiratorypressuresinhealthyyoungadults