Osteolysis following PE Wear of a Hastings Head on a Monoblock Hip Stem

We report a case of extended osteolysis, requiring a third revision of the left hip in an 85-year-old man 46 years after index operation. Major polyethylene (PE) wear occurred due to a missmatched combination of a bipolar Hastings head with a PE liner and head damage of the originally maintained ste...

Full description

Saved in:
Bibliographic Details
Main Authors: Sarah Fischer, Valerie Polster, Miriam Ruhr, Robert Hube, Michael M. Morlock
Format: Article
Language:English
Published: Wiley 2021-01-01
Series:Case Reports in Orthopedics
Online Access:http://dx.doi.org/10.1155/2021/9989395
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report a case of extended osteolysis, requiring a third revision of the left hip in an 85-year-old man 46 years after index operation. Major polyethylene (PE) wear occurred due to a missmatched combination of a bipolar Hastings head with a PE liner and head damage of the originally maintained stem. This case demonstrates that bipolar heads should not be used with PE cup liners since the respective bearing diameters cannot be guaranteed to match due to missing specifications. Furthermore, putting a Hastings head on an already damaged head of the stem should be omitted and rather the stem should initially be revised.