Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability

Purpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete ant...

Full description

Saved in:
Bibliographic Details
Main Authors: David P. Piñero, Alberto López-Navarro, Inmaculada Cabezos, Dolores de Fez, María T. Caballero, Vicent J. Camps
Format: Article
Language:English
Published: Wiley 2017-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2017/2149145
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832563049075048448
author David P. Piñero
Alberto López-Navarro
Inmaculada Cabezos
Dolores de Fez
María T. Caballero
Vicent J. Camps
author_facet David P. Piñero
Alberto López-Navarro
Inmaculada Cabezos
Dolores de Fez
María T. Caballero
Vicent J. Camps
author_sort David P. Piñero
collection DOAJ
description Purpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete anterior segment examination with the VX120 system (Visionix-Luneau Technologies, Chartres, France). Three consecutive measurements were obtained. The within-subject standard deviation (Sw), intrasubject precision (1.96×Sw), and intraclass correlation coefficient (ICC) were calculated. Results. All Sw for corneal power measurements were below 0.26 D, with ICC above 0.982. The Sw for corneal astigmatism at different areas (3, 5, and 7 mm) was below 0.21 D, with ICC above 0.913. Concerning the axis of astigmatism, its Sw was below 11.27°, with ICC above 0.975. The Sw and ICC for corneal eccentricity were 0.067 and 0.957, respectively. The Sw and ICC for high-order aberration root mean square (RMS) were 0.048 µm and 0.901, respectively. For 3rd- and 4th-order aberrometric parameters, all Sw were below 0.037 µm and all ICC were higher than 0.84, except for quadrafoil RMS (ICC: 0.689). Conclusions. The multidiagnostic device evaluated is able to provide consistent measurements of corneal power, eccentricity, and third- and fourth-order aberrations in healthy eyes.
format Article
id doaj-art-bed9fe8a589946379e3640fda55bcac2
institution Kabale University
issn 2090-004X
2090-0058
language English
publishDate 2017-01-01
publisher Wiley
record_format Article
series Journal of Ophthalmology
spelling doaj-art-bed9fe8a589946379e3640fda55bcac22025-02-03T01:21:10ZengWileyJournal of Ophthalmology2090-004X2090-00582017-01-01201710.1155/2017/21491452149145Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession RepeatabilityDavid P. Piñero0Alberto López-Navarro1Inmaculada Cabezos2Dolores de Fez3María T. Caballero4Vicent J. Camps5Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainOptometry Clinic, Prevention Service of the University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainPurpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete anterior segment examination with the VX120 system (Visionix-Luneau Technologies, Chartres, France). Three consecutive measurements were obtained. The within-subject standard deviation (Sw), intrasubject precision (1.96×Sw), and intraclass correlation coefficient (ICC) were calculated. Results. All Sw for corneal power measurements were below 0.26 D, with ICC above 0.982. The Sw for corneal astigmatism at different areas (3, 5, and 7 mm) was below 0.21 D, with ICC above 0.913. Concerning the axis of astigmatism, its Sw was below 11.27°, with ICC above 0.975. The Sw and ICC for corneal eccentricity were 0.067 and 0.957, respectively. The Sw and ICC for high-order aberration root mean square (RMS) were 0.048 µm and 0.901, respectively. For 3rd- and 4th-order aberrometric parameters, all Sw were below 0.037 µm and all ICC were higher than 0.84, except for quadrafoil RMS (ICC: 0.689). Conclusions. The multidiagnostic device evaluated is able to provide consistent measurements of corneal power, eccentricity, and third- and fourth-order aberrations in healthy eyes.http://dx.doi.org/10.1155/2017/2149145
spellingShingle David P. Piñero
Alberto López-Navarro
Inmaculada Cabezos
Dolores de Fez
María T. Caballero
Vicent J. Camps
Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
Journal of Ophthalmology
title Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
title_full Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
title_fullStr Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
title_full_unstemmed Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
title_short Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
title_sort corneal topographic and aberrometric measurements obtained with a multidiagnostic device in healthy eyes intrasession repeatability
url http://dx.doi.org/10.1155/2017/2149145
work_keys_str_mv AT davidppinero cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability
AT albertolopeznavarro cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability
AT inmaculadacabezos cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability
AT doloresdefez cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability
AT mariatcaballero cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability
AT vicentjcamps cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability