Continuous Monitoring of Vital Signs After Hospital Discharge: A Feasibility Study
# Introduction Increasing demand for inpatient beds limits capacity and poses a challenge to the healthcare system. Early discharge may be one solution to solve this problem, and continuous vital sign monitoring at home could safely facilitate this goal. We aimed to document feasibility of continuo...
Saved in:
| Main Authors: | , , , , , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Patient Safety Authority
2023-06-01
|
| Series: | Patient Safety |
| Online Access: | https://doi.org/10.33940/001c.77776 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | # Introduction
Increasing demand for inpatient beds limits capacity and poses a challenge to the healthcare system. Early discharge may be one solution to solve this problem, and continuous vital sign monitoring at home could safely facilitate this goal. We aimed to document feasibility of continuous home monitoring in patients after hospital discharge.
# Methods
Patients were eligible for inclusion if they were admitted with acute medical disease and scheduled for discharge. They wore three wireless vital sign sensors for four days at home: a chest patch measuring heart rate and respiratory rate, a pulse oximeter, and a blood pressure (BP) monitor. Patients with ≥6 hours monitoring time after discharge were included in the analysis. Primary outcome was percentage of maximum monitoring time of heart rate and respiratory rate.
# Results
Monitoring was initiated in 80 patients, and 69 patients (86%) had ≥6 hours monitoring time after discharge. The chest patch, pulse oximeter, and BP monitor collected data for 88%, 60%, and 32% of the monitored time, respectively. Oxygen desaturation \<88% was observed in 92% of the patients and lasted for 6.3% (interquartile range \[IQR\] 0.9%--22.0%) of total monitoring time. Desaturation below 85% was observed in 83% of the patients and lasted 4.2% \[IQR 0.4%--9.4%\] of total monitoring time. 61% had tachypnea (\>24/minute); tachycardia (\>130/minute) lasting ≥30 minutes was observed in 28% of the patients.
# Conclusions
Continuous monitoring of vital signs was feasible at home with a high degree of valid monitoring time. Oxygen desaturation was commonly observed. |
|---|---|