Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.

The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with 177Lu...

Full description

Saved in:
Bibliographic Details
Main Authors: Emilio Mezzenga, Vincenzo D'Errico, Marco D'Arienzo, Lidia Strigari, Koutla Panagiota, Federica Matteucci, Stefano Severi, Giovanni Paganelli, Andrew Fenwick, David Bianchini, Francesco Marcocci, Anna Sarnelli
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2017-01-01
Series:PLoS ONE
Online Access:https://journals.plos.org/plosone/article/file?id=10.1371/journal.pone.0182888&type=printable
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850229080389058560
author Emilio Mezzenga
Vincenzo D'Errico
Marco D'Arienzo
Lidia Strigari
Koutla Panagiota
Federica Matteucci
Stefano Severi
Giovanni Paganelli
Andrew Fenwick
David Bianchini
Francesco Marcocci
Anna Sarnelli
author_facet Emilio Mezzenga
Vincenzo D'Errico
Marco D'Arienzo
Lidia Strigari
Koutla Panagiota
Federica Matteucci
Stefano Severi
Giovanni Paganelli
Andrew Fenwick
David Bianchini
Francesco Marcocci
Anna Sarnelli
author_sort Emilio Mezzenga
collection DOAJ
description The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with 177Lu. The accuracy of 177Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.
format Article
id doaj-art-ae68a6ae3fac48188f46bbb50c65af39
institution OA Journals
issn 1932-6203
language English
publishDate 2017-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj-art-ae68a6ae3fac48188f46bbb50c65af392025-08-20T02:04:20ZengPublic Library of Science (PLoS)PLoS ONE1932-62032017-01-01128e018288810.1371/journal.pone.0182888Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.Emilio MezzengaVincenzo D'ErricoMarco D'ArienzoLidia StrigariKoutla PanagiotaFederica MatteucciStefano SeveriGiovanni PaganelliAndrew FenwickDavid BianchiniFrancesco MarcocciAnna SarnelliThe purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with 177Lu. The accuracy of 177Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.https://journals.plos.org/plosone/article/file?id=10.1371/journal.pone.0182888&type=printable
spellingShingle Emilio Mezzenga
Vincenzo D'Errico
Marco D'Arienzo
Lidia Strigari
Koutla Panagiota
Federica Matteucci
Stefano Severi
Giovanni Paganelli
Andrew Fenwick
David Bianchini
Francesco Marcocci
Anna Sarnelli
Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.
PLoS ONE
title Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.
title_full Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.
title_fullStr Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.
title_full_unstemmed Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.
title_short Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.
title_sort quantitative accuracy of 177lu spect imaging for molecular radiotherapy
url https://journals.plos.org/plosone/article/file?id=10.1371/journal.pone.0182888&type=printable
work_keys_str_mv AT emiliomezzenga quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT vincenzoderrico quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT marcodarienzo quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT lidiastrigari quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT koutlapanagiota quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT federicamatteucci quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT stefanoseveri quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT giovannipaganelli quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT andrewfenwick quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT davidbianchini quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT francescomarcocci quantitativeaccuracyof177luspectimagingformolecularradiotherapy
AT annasarnelli quantitativeaccuracyof177luspectimagingformolecularradiotherapy