Fracture of a Femoral Revision Stem following a Technical Failure

We report about a fracture of a modular, uncemented femoral revision stem (Zimmer, Revitan®) due to a not previously described intraoperative technical problem. During implantation, a small ring, part of the proximal part of the trial stem, was left on the distal part of the definite stem. Following...

Full description

Saved in:
Bibliographic Details
Main Authors: Falko Herold, Henk Eijer
Format: Article
Language:English
Published: Wiley 2018-01-01
Series:Case Reports in Orthopedics
Online Access:http://dx.doi.org/10.1155/2018/9691627
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report about a fracture of a modular, uncemented femoral revision stem (Zimmer, Revitan®) due to a not previously described intraoperative technical problem. During implantation, a small ring, part of the proximal part of the trial stem, was left on the distal part of the definite stem. Following this, the top screwcap of the proximal part of the definite stem could not be tightened properly. However, the stem was thought to be stable, and the situation left. Two and a half years later, the proximal part of the stem fractured and the situation became unstable. It is very useful to know about this ring and that it should always be removed, otherwise, there is possibility that it may be left on the definite distal part of the stem with the possibility of a later fracture.