Determination of hexachlorocyclopentadiene in water by HS-SPME-GC-MS

A headspace solid-phase microextraction gas chromatography-mass spectrometry(HS-SPME-GC-MS) method was established for the determination of hexachlorocyclopentadiene in water, with using headspace solid-phase microextraction injection, DB-1 chromatographic column separation, and gas chromatography/m...

Full description

Saved in:
Bibliographic Details
Main Authors: YOU Xihua, SHI Jinghua, WANG Weili, CAO Fangfang, ZHANG Fengju, QIN Chenghua
Format: Article
Language:zho
Published: Editorial Office of Industrial Water Treatment 2025-01-01
Series:Gongye shui chuli
Subjects:
Online Access:https://www.iwt.cn/CN/10.19965/j.cnki.iwt.2023-1226
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A headspace solid-phase microextraction gas chromatography-mass spectrometry(HS-SPME-GC-MS) method was established for the determination of hexachlorocyclopentadiene in water, with using headspace solid-phase microextraction injection, DB-1 chromatographic column separation, and gas chromatography/mass spectrometry detector detection. The effects of coating solid-phase extraction needles, inlet temperatures, solid-phase extraction conditions, salt effects on the determination were systematically investigated, and the method was validated under the optimal experimental conditions. The results showed that hexachlorocyclopentadiene exhibited good linearity with chromatographic peak area within the mass concentration range of 0.1-10 µg/L, with a correlation coefficient of 0.997 and a detection limit of 0.03 µg/L. The actual surface water and wastewater samples were measured, and the monitoring result of wastewater sample was 0.5 µg/L, while the monitoring result of surface water sample was not detected. The actual samples were spiked with 0.5 µg/L and 1.0 µg/L, respectively. The recovery rate was 70%-87%, and the relative standard deviation was 4.14%-13.9%(n=6). This method was simple to operate, did not use organic solvents for extraction, had good precision and high accuracy, which could be used for rapid determination of hexachlorocyclopentadiene in industrial wastewater and surface water.
ISSN:1005-829X