How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases
Abstract Background With 3D printing technology, we can now use preoperative imaging for precise surgical plan. We can also use patient-specific surgical jig to improve the accuracy of osteotomy and 3D-printed custom-made endoprostheses combined with a screw-rod system to restore lumbosacral stabili...
Saved in:
| Main Authors: | , , , , , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
BMC
2025-07-01
|
| Series: | 3D Printing in Medicine |
| Subjects: | |
| Online Access: | https://doi.org/10.1186/s41205-025-00295-6 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| _version_ | 1849767157812953088 |
|---|---|
| author | Qianyu Shi Jiazhi Zhu Haijie Liang Ruifeng Wang Siyi Huang Wei Guo Tao Ji Xiaodong Tang |
| author_facet | Qianyu Shi Jiazhi Zhu Haijie Liang Ruifeng Wang Siyi Huang Wei Guo Tao Ji Xiaodong Tang |
| author_sort | Qianyu Shi |
| collection | DOAJ |
| description | Abstract Background With 3D printing technology, we can now use preoperative imaging for precise surgical plan. We can also use patient-specific surgical jig to improve the accuracy of osteotomy and 3D-printed custom-made endoprostheses combined with a screw-rod system to restore lumbosacral stability. The aim of this study was to evaluate the accuracy of 3D printing technology for precise osteotomy during total sacrectomy. Methods Nine patients with primary malignant tumors of the sacrum who underwent total sacrectomy at our center were enrolled. Osteotomy was planned based on preoperative imaging (CT, MRI). Generally, an additional 8–10 mm margin beyond the tumor was determined by the fusion of MR and CT images. Patient-specific surgical jigs and 3D-printed sacral endoprostheses were then designed based on the planned osteotomy planes. Pre- and postoperative 3D models of the lumbosacral and pelvic regions were constructed using the fiducial registration model of 3D slicer software 5.1.0. Postoperative CT scans were compared with the planned osteomy planes based on preoperative CT scans, in order to evaluate the accuracy of the osteotomy and endoprosthetic reconstruction. For each patient, four levels of osteotomy planes were chosen, including the upper edge of the sacroiliac (SI) joint, the S1 and S2 foramen levels, and the caudal edge of the SI joint, for analyzing position and angular deviations between the preoperative plan and actual osteotomy along with the endoprosthesis position. Results Pathological diagnoses included four cases of osteosarcoma, four cases of chordoma, and one case of Ewing sarcoma. All osteotomies in nine patients achieved R0 resection, as verified pathologically. An average angular deviation of 4.27° (interquartile range[IQR] 4.15) and an osteotomy position deviation of 4.00 mm (IQR 2.90) were observed. The mean angular deviations of the four levels were 3.50° (IQR 6.02), 3.86° (IQR 2.55), 4.81° (IQR 4.37), and 4.92° (IQR 3.27). The mean position deviations at the four levels were 3.15 mm (IQR 3.54), 3.55 mm (IQR 1.37), 4.26 mm (IQR 2.61), and 4.86 mm (IQR 3.93). No significant difference was found among the angular and position deviations at different levels. However, the proportions of individuals with position deviations > 2 mm and > 5 mm were significantly greater at the caudal end of the SI joint than at the upper end. All position deviations were within 8 mm. The average follow-up duration was 24.4 months. At the last follow-up, three patients experienced local recurrence, and one patient died of disease. All endoprostheses were in place without significant displacement. The mean Musculoskeletal Tumor Society scoring system (MSTS93) and MUD scores (function and sensation of lower limbs (M), urination and uriesthesia (U), and defecation and rectal sensation (D)) were 19.4 (16 to 24) and 16.3 (12 to 24), respectively. Conclusion Notably, 3D-printed patient-specific surgical jigs exhibit high accuracy of osteotomy and lead to optimal surgical margin and reconstruction in total sacrectomy. Effective and reliable reconstruction can be achieved with a custom-made 3D-printed endoprosthesis. The application of 3D printing technology using patient-specific surgical jigs and the custom-made 3D-printed implants exhibited high surgical accuracy in total sacrectomy, as evidenced by accuracy validation. |
| format | Article |
| id | doaj-art-a86aa90223714bc89c2b7863df1ad1e0 |
| institution | DOAJ |
| issn | 2365-6271 |
| language | English |
| publishDate | 2025-07-01 |
| publisher | BMC |
| record_format | Article |
| series | 3D Printing in Medicine |
| spelling | doaj-art-a86aa90223714bc89c2b7863df1ad1e02025-08-20T03:04:18ZengBMC3D Printing in Medicine2365-62712025-07-0111111210.1186/s41205-025-00295-6How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive casesQianyu Shi0Jiazhi Zhu1Haijie Liang2Ruifeng Wang3Siyi Huang4Wei Guo5Tao Ji6Xiaodong Tang7Department of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityDepartment of Musculoskeletal Tumor, People’s Hospital, Peking UniversityAbstract Background With 3D printing technology, we can now use preoperative imaging for precise surgical plan. We can also use patient-specific surgical jig to improve the accuracy of osteotomy and 3D-printed custom-made endoprostheses combined with a screw-rod system to restore lumbosacral stability. The aim of this study was to evaluate the accuracy of 3D printing technology for precise osteotomy during total sacrectomy. Methods Nine patients with primary malignant tumors of the sacrum who underwent total sacrectomy at our center were enrolled. Osteotomy was planned based on preoperative imaging (CT, MRI). Generally, an additional 8–10 mm margin beyond the tumor was determined by the fusion of MR and CT images. Patient-specific surgical jigs and 3D-printed sacral endoprostheses were then designed based on the planned osteotomy planes. Pre- and postoperative 3D models of the lumbosacral and pelvic regions were constructed using the fiducial registration model of 3D slicer software 5.1.0. Postoperative CT scans were compared with the planned osteomy planes based on preoperative CT scans, in order to evaluate the accuracy of the osteotomy and endoprosthetic reconstruction. For each patient, four levels of osteotomy planes were chosen, including the upper edge of the sacroiliac (SI) joint, the S1 and S2 foramen levels, and the caudal edge of the SI joint, for analyzing position and angular deviations between the preoperative plan and actual osteotomy along with the endoprosthesis position. Results Pathological diagnoses included four cases of osteosarcoma, four cases of chordoma, and one case of Ewing sarcoma. All osteotomies in nine patients achieved R0 resection, as verified pathologically. An average angular deviation of 4.27° (interquartile range[IQR] 4.15) and an osteotomy position deviation of 4.00 mm (IQR 2.90) were observed. The mean angular deviations of the four levels were 3.50° (IQR 6.02), 3.86° (IQR 2.55), 4.81° (IQR 4.37), and 4.92° (IQR 3.27). The mean position deviations at the four levels were 3.15 mm (IQR 3.54), 3.55 mm (IQR 1.37), 4.26 mm (IQR 2.61), and 4.86 mm (IQR 3.93). No significant difference was found among the angular and position deviations at different levels. However, the proportions of individuals with position deviations > 2 mm and > 5 mm were significantly greater at the caudal end of the SI joint than at the upper end. All position deviations were within 8 mm. The average follow-up duration was 24.4 months. At the last follow-up, three patients experienced local recurrence, and one patient died of disease. All endoprostheses were in place without significant displacement. The mean Musculoskeletal Tumor Society scoring system (MSTS93) and MUD scores (function and sensation of lower limbs (M), urination and uriesthesia (U), and defecation and rectal sensation (D)) were 19.4 (16 to 24) and 16.3 (12 to 24), respectively. Conclusion Notably, 3D-printed patient-specific surgical jigs exhibit high accuracy of osteotomy and lead to optimal surgical margin and reconstruction in total sacrectomy. Effective and reliable reconstruction can be achieved with a custom-made 3D-printed endoprosthesis. The application of 3D printing technology using patient-specific surgical jigs and the custom-made 3D-printed implants exhibited high surgical accuracy in total sacrectomy, as evidenced by accuracy validation.https://doi.org/10.1186/s41205-025-00295-6Precise excision and reconstructionTotal sacrectomySurgical accuracy validation3D printing technologyPatient-specific instrument |
| spellingShingle | Qianyu Shi Jiazhi Zhu Haijie Liang Ruifeng Wang Siyi Huang Wei Guo Tao Ji Xiaodong Tang How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases 3D Printing in Medicine Precise excision and reconstruction Total sacrectomy Surgical accuracy validation 3D printing technology Patient-specific instrument |
| title | How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases |
| title_full | How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases |
| title_fullStr | How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases |
| title_full_unstemmed | How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases |
| title_short | How precise is excision and reconstruction using 3D printing technology for total sacrectomy: accuracy validation in 9 consecutive cases |
| title_sort | how precise is excision and reconstruction using 3d printing technology for total sacrectomy accuracy validation in 9 consecutive cases |
| topic | Precise excision and reconstruction Total sacrectomy Surgical accuracy validation 3D printing technology Patient-specific instrument |
| url | https://doi.org/10.1186/s41205-025-00295-6 |
| work_keys_str_mv | AT qianyushi howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT jiazhizhu howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT haijieliang howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT ruifengwang howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT siyihuang howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT weiguo howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT taoji howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases AT xiaodongtang howpreciseisexcisionandreconstructionusing3dprintingtechnologyfortotalsacrectomyaccuracyvalidationin9consecutivecases |