On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field

Sources of pulsed photon radiation are widely used in science and industry. In this regard, there is a growing demand for dosimetry instruments capable of operating in these pulsed photon radiation fields. Until now there is no comprehensive methodology that describes the characteristics of referenc...

Full description

Saved in:
Bibliographic Details
Main Authors: A. A. Zaharadniuk, A. Y. Taraev, S. V. Lazarenko
Format: Article
Language:English
Published: Belarusian National Technical University 2023-10-01
Series:Приборы и методы измерений
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/830
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849687683506372608
author A. A. Zaharadniuk
A. Y. Taraev
S. V. Lazarenko
author_facet A. A. Zaharadniuk
A. Y. Taraev
S. V. Lazarenko
author_sort A. A. Zaharadniuk
collection DOAJ
description Sources of pulsed photon radiation are widely used in science and industry. In this regard, there is a growing demand for dosimetry instruments capable of operating in these pulsed photon radiation fields. Until now there is no comprehensive methodology that describes the characteristics of reference fields of pulsed photon radiation required for calibration and verification of dosimetry equipment. The aim of this paper was to study the possibility of using a medical linear electron accelerator as a generating source of a reference pulsed photon radiation field. The paper investigates main characteristics of photon radiation field (namely, spectrum and average energy) generated by CLINAC in two modes of operation: 6 MV and 18 MV. Additionally it researches the possibility of attenuation of the CLINAC photon radiation intensity by a lead filter. The spectrum and average energy of the CLINAC photon radiation were calculated by means of Monte Carlo simulation in the Fluka program. The validation of the calculation was determined by comparing the attenuation coefficients calculated by the model with ones obtained experimentally by using real CLINACs. Experimentally, the attenuation coefficients were obtained by measuring the air kerma rate generated by the photon radiation fields of Varian VitalBeam and Varian iX CLINACs at a given point of space. Based on the simulation results, the Ḣ*(10) ambient dose equivalent ratio generated by CLINACs were calculated. It was found that the lead filter effectively attenuates the CLINAC photon radiation in terms of both air kerma rate and ambient dose equivalent rate to levels suitable for calibration of dosimetry equipment designed to measure pulsed radiation. It is shown that the lead filter significantly affects both the photon spectrum of the CLINAC and its average energy.
format Article
id doaj-art-a75c398b0bfa455799164c8f2a2989cf
institution DOAJ
issn 2220-9506
2414-0473
language English
publishDate 2023-10-01
publisher Belarusian National Technical University
record_format Article
series Приборы и методы измерений
spelling doaj-art-a75c398b0bfa455799164c8f2a2989cf2025-08-20T03:22:15ZengBelarusian National Technical UniversityПриборы и методы измерений2220-95062414-04732023-10-0114317919010.21122/2220-9506-2023-14-3-179-190621On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation FieldA. A. Zaharadniuk0A. Y. Taraev1S. V. Lazarenko2ATOMTEXATOMTEXATOMTEXSources of pulsed photon radiation are widely used in science and industry. In this regard, there is a growing demand for dosimetry instruments capable of operating in these pulsed photon radiation fields. Until now there is no comprehensive methodology that describes the characteristics of reference fields of pulsed photon radiation required for calibration and verification of dosimetry equipment. The aim of this paper was to study the possibility of using a medical linear electron accelerator as a generating source of a reference pulsed photon radiation field. The paper investigates main characteristics of photon radiation field (namely, spectrum and average energy) generated by CLINAC in two modes of operation: 6 MV and 18 MV. Additionally it researches the possibility of attenuation of the CLINAC photon radiation intensity by a lead filter. The spectrum and average energy of the CLINAC photon radiation were calculated by means of Monte Carlo simulation in the Fluka program. The validation of the calculation was determined by comparing the attenuation coefficients calculated by the model with ones obtained experimentally by using real CLINACs. Experimentally, the attenuation coefficients were obtained by measuring the air kerma rate generated by the photon radiation fields of Varian VitalBeam and Varian iX CLINACs at a given point of space. Based on the simulation results, the Ḣ*(10) ambient dose equivalent ratio generated by CLINACs were calculated. It was found that the lead filter effectively attenuates the CLINAC photon radiation in terms of both air kerma rate and ambient dose equivalent rate to levels suitable for calibration of dosimetry equipment designed to measure pulsed radiation. It is shown that the lead filter significantly affects both the photon spectrum of the CLINAC and its average energy.https://pimi.bntu.by/jour/article/view/830monte carlo modellinglinear electron acceleratorpulsed photon radiation
spellingShingle A. A. Zaharadniuk
A. Y. Taraev
S. V. Lazarenko
On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field
Приборы и методы измерений
monte carlo modelling
linear electron accelerator
pulsed photon radiation
title On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field
title_full On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field
title_fullStr On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field
title_full_unstemmed On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field
title_short On the Possibility of Using Medical Linear Electron Accelerators as a Source of the Reference Pulsed Photon Radiation Field
title_sort on the possibility of using medical linear electron accelerators as a source of the reference pulsed photon radiation field
topic monte carlo modelling
linear electron accelerator
pulsed photon radiation
url https://pimi.bntu.by/jour/article/view/830
work_keys_str_mv AT aazaharadniuk onthepossibilityofusingmedicallinearelectronacceleratorsasasourceofthereferencepulsedphotonradiationfield
AT aytaraev onthepossibilityofusingmedicallinearelectronacceleratorsasasourceofthereferencepulsedphotonradiationfield
AT svlazarenko onthepossibilityofusingmedicallinearelectronacceleratorsasasourceofthereferencepulsedphotonradiationfield