The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2

Objectives: This study aimed to evaluate the potential relationship between the time threshold (Tt) values of a commercial EasyNAT system, which is based on cross priming amplification (CPA) technology, and the cycle threshold (Ct) values of the Cobas z480 analyzer, which is based on a real-time flu...

Full description

Saved in:
Bibliographic Details
Main Authors: Lu-Qing Zheng, Qing-Yong Wang
Format: Article
Language:English
Published: Elsevier 2024-11-01
Series:Practical Laboratory Medicine
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2352551724000775
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850251267214934016
author Lu-Qing Zheng
Qing-Yong Wang
author_facet Lu-Qing Zheng
Qing-Yong Wang
author_sort Lu-Qing Zheng
collection DOAJ
description Objectives: This study aimed to evaluate the potential relationship between the time threshold (Tt) values of a commercial EasyNAT system, which is based on cross priming amplification (CPA) technology, and the cycle threshold (Ct) values of the Cobas z480 analyzer, which is based on a real-time fluorescence polymerase chain reaction (PCR) method, in the detection of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) from oropharyngeal swabs. Design and Methods: Data were retrospectively collected from a clinical laboratory between December 4, 2022 and July 1, 2024. Results: A total of 277 EasyNAT-positive samples (Tt values from 3.83 to 29.5) were simultaneously investigated using the Cobas z480 analyzer (Ct values from 10.74 to 38.78). The concordance rate between the two systems was 100 %. Among the positive samples, the mean and maximum PCR Ct values of O and N genes increased in line with increasing Tt values of the left and right amplification areas of the EasyNAT system. The maximum Ct values of the O or N gene determined by the Cobas z480 analyzer were no more than 29.52 when the Tt values of the left or right amplification areas of the UC0116 analyzer were no more than 6. Conclusions: The safe, simple, fast, accurate, and automatic EasyNAT system used in conjunction with a PCR system might be a better choice for the detection of SARS-CoV-2 in hospitals, especially in settings without sophisticated PCR facilities. The Tt value (≤6) of the EasyNAT system can be a reference index for estimating the maximum Ct value (29.52) in SARS-CoV-2-positive samples.
format Article
id doaj-art-a2402ca61bee44caba22fa9827541984
institution OA Journals
issn 2352-5517
language English
publishDate 2024-11-01
publisher Elsevier
record_format Article
series Practical Laboratory Medicine
spelling doaj-art-a2402ca61bee44caba22fa98275419842025-08-20T01:57:56ZengElsevierPractical Laboratory Medicine2352-55172024-11-0142e0043110.1016/j.plabm.2024.e00431The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2Lu-Qing Zheng0Qing-Yong Wang1Department of Clinical Laboratory, Xiyuan Hospital, China Academy of Chinese Medical Sciences, Beijing, 100091, ChinaCorresponding author. Department of Clinical Laboratory, Xiyuan Hospital, China Academy of Chinese Medical Sciences, Beijing, 100091, China.; Department of Clinical Laboratory, Xiyuan Hospital, China Academy of Chinese Medical Sciences, Beijing, 100091, ChinaObjectives: This study aimed to evaluate the potential relationship between the time threshold (Tt) values of a commercial EasyNAT system, which is based on cross priming amplification (CPA) technology, and the cycle threshold (Ct) values of the Cobas z480 analyzer, which is based on a real-time fluorescence polymerase chain reaction (PCR) method, in the detection of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) from oropharyngeal swabs. Design and Methods: Data were retrospectively collected from a clinical laboratory between December 4, 2022 and July 1, 2024. Results: A total of 277 EasyNAT-positive samples (Tt values from 3.83 to 29.5) were simultaneously investigated using the Cobas z480 analyzer (Ct values from 10.74 to 38.78). The concordance rate between the two systems was 100 %. Among the positive samples, the mean and maximum PCR Ct values of O and N genes increased in line with increasing Tt values of the left and right amplification areas of the EasyNAT system. The maximum Ct values of the O or N gene determined by the Cobas z480 analyzer were no more than 29.52 when the Tt values of the left or right amplification areas of the UC0116 analyzer were no more than 6. Conclusions: The safe, simple, fast, accurate, and automatic EasyNAT system used in conjunction with a PCR system might be a better choice for the detection of SARS-CoV-2 in hospitals, especially in settings without sophisticated PCR facilities. The Tt value (≤6) of the EasyNAT system can be a reference index for estimating the maximum Ct value (29.52) in SARS-CoV-2-positive samples.http://www.sciencedirect.com/science/article/pii/S2352551724000775EasyNAT systemCobas z480Time thresholdCycle thresholdSARS-CoV-2
spellingShingle Lu-Qing Zheng
Qing-Yong Wang
The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2
Practical Laboratory Medicine
EasyNAT system
Cobas z480
Time threshold
Cycle threshold
SARS-CoV-2
title The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2
title_full The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2
title_fullStr The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2
title_full_unstemmed The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2
title_short The potential relationship between EasyNAT system Tt values and Cobas z480 Ct values in the detection of SARS-Cov-2
title_sort potential relationship between easynat system tt values and cobas z480 ct values in the detection of sars cov 2
topic EasyNAT system
Cobas z480
Time threshold
Cycle threshold
SARS-CoV-2
url http://www.sciencedirect.com/science/article/pii/S2352551724000775
work_keys_str_mv AT luqingzheng thepotentialrelationshipbetweeneasynatsystemttvaluesandcobasz480ctvaluesinthedetectionofsarscov2
AT qingyongwang thepotentialrelationshipbetweeneasynatsystemttvaluesandcobasz480ctvaluesinthedetectionofsarscov2
AT luqingzheng potentialrelationshipbetweeneasynatsystemttvaluesandcobasz480ctvaluesinthedetectionofsarscov2
AT qingyongwang potentialrelationshipbetweeneasynatsystemttvaluesandcobasz480ctvaluesinthedetectionofsarscov2