The Optimization of a T-Cell Resonator: Towards Highly Sensitive Photoacoustic Spectroscopy for Noninvasive Blood Glucose Detection
Noninvasive blood glucose monitoring is crucial for diabetes management, and photoacoustic spectroscopy (PAS) offers a promising solution by detecting glucose levels through human skin. However, weak acoustic signals in PAS systems require optimized resonator designs for enhanced detection sensitivi...
Saved in:
| Main Authors: | , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
MDPI AG
2025-04-01
|
| Series: | Biosensors |
| Subjects: | |
| Online Access: | https://www.mdpi.com/2079-6374/15/4/254 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | Noninvasive blood glucose monitoring is crucial for diabetes management, and photoacoustic spectroscopy (PAS) offers a promising solution by detecting glucose levels through human skin. However, weak acoustic signals in PAS systems require optimized resonator designs for enhanced detection sensitivity. Designing such resonators physically is complex, requiring the precise identification of critical parameters before practical implementation. This study focused on optimizing a T-shaped photoacoustic resonator using finite element modeling in a COMSOL Multiphysics environment. By systematically varying the geometric design parameters of the T-cell resonator, a maximum increase in the pressure amplitude of 12.76 times with a quality factor (Q-factor) of 47.5 was achieved compared to the previously designed reference acoustic resonator. This study took a significant step forward by identifying key geometric parameters that influence resonator performance, paving the way for more sensitive and reliable noninvasive glucose monitoring systems. |
|---|---|
| ISSN: | 2079-6374 |