Elimination of artifacts caused by residual radiopharmaceutical activity in injection site in myocardial perfusion imaging on Discovery NM 530c semiconductor gamma camera
Abstract Background Cardiac gamma cameras dedicated for myocardial perfusion imaging (MPI) perform studies faster and acquire higher quality images than traditional cameras. However, they are susceptible to some artifacts. We observed a previously unreported artifact, caused by residual radiopharmac...