Modification of Screen-Printed Carbon Electrode (SPCE) by Magnetic Fe3O4-Chitosan for Detection of Hexavalent Chromium

A new modification of screen-printed carbon electrode (SPCE) with magnetic Fe3O4-chitosan has been developed. Magnetic Fe3O4-chitosan was made from a mixture of Fe2+/Fe3+ (1:2) with 1% chitosan. Sodium tripolyphosphate was used as a crosslinking reagent for chitosan under alkaline conditions using N...

Full description

Saved in:
Bibliographic Details
Main Authors: Ani Mulyasuryani, Yuniar Ponco Prananto
Format: Article
Language:English
Published: Universitas Islam Indonesia 2025-03-01
Series:Indonesian Journal of Chemical Analysis
Subjects:
Online Access:https://journal.uii.ac.id/IJCA/article/view/39263
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A new modification of screen-printed carbon electrode (SPCE) with magnetic Fe3O4-chitosan has been developed. Magnetic Fe3O4-chitosan was made from a mixture of Fe2+/Fe3+ (1:2) with 1% chitosan. Sodium tripolyphosphate was used as a crosslinking reagent for chitosan under alkaline conditions using NH3. SPCE modification with Fe3O4 can increase the sensitivity of Cr(VI) detection by differential pulse voltammetry (DPV). This method resulted in a sensitivity of 21.9 ppm/μA in the 0.1 - 1 ppm Cr(VI) concentration range in a 0.005 M KCl-HCl electrolyte mixture. The detection limit (LoD) is 0.2 ppm with an average accuracy of 94%.
ISSN:2622-7401
2622-7126