METHODS AND HARDWARE OF DOSE OUTPUT VERIFICATION FOR DYNAMIC RADIOTHERAPY

The design of special verification phantom for dynamic radiotherapy checking is described. This phantom permits to insert the dose distribution cross-calibration before every days patients irradiation on Linac with RapidArc. Cross-calibration factor is defined by approximation of large number correc...

Full description

Saved in:
Bibliographic Details
Main Authors: Y. V. Tsitovich, A. I. Hmyrak, A. I. Tarutin, M. G. Kiselev
Format: Article
Language:English
Published: Belarusian National Technical University 2015-03-01
Series:Приборы и методы измерений
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/24
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The design of special verification phantom for dynamic radiotherapy checking is described. This phantom permits to insert the dose distribution cross-calibration before every days patients irradiation on Linac with RapidArc. Cross-calibration factor is defined by approximation of large number correction factors measured in phantom at different angles of gantry rotation and middle quantity calculation. The long range stability of all correction factors have been evaluated during checking of setting treatment plans and did not exceed 1 %. The fast checking of patient’s dynamic irradiation with developed method gives possibility texactly realize of planning dose distribution and to increase the capacity of accelerators.