Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism?
Unexplained change in parameters, despite good lead positions or odd ECG patterns following implant should lead to suspicion of lead swap, amongst other possibilities. An easy way to confirm this is by paying careful attention to device EGMs and recording the ECG by switching off one channel, and pa...
Saved in:
| Main Authors: | , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Elsevier
2025-01-01
|
| Series: | Indian Pacing and Electrophysiology Journal |
| Subjects: | |
| Online Access: | http://www.sciencedirect.com/science/article/pii/S0972629224001529 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| _version_ | 1850069272183701504 |
|---|---|
| author | Arpita Katheria Kamlesh Raut Ankit Kumar Sahu Aditya Kapoor |
| author_facet | Arpita Katheria Kamlesh Raut Ankit Kumar Sahu Aditya Kapoor |
| author_sort | Arpita Katheria |
| collection | DOAJ |
| description | Unexplained change in parameters, despite good lead positions or odd ECG patterns following implant should lead to suspicion of lead swap, amongst other possibilities. An easy way to confirm this is by paying careful attention to device EGMs and recording the ECG by switching off one channel, and pacing single channel (in either AAI or VVI mode). |
| format | Article |
| id | doaj-art-8df69042273e456fb7cc2c7db3bd4e9b |
| institution | DOAJ |
| issn | 0972-6292 |
| language | English |
| publishDate | 2025-01-01 |
| publisher | Elsevier |
| record_format | Article |
| series | Indian Pacing and Electrophysiology Journal |
| spelling | doaj-art-8df69042273e456fb7cc2c7db3bd4e9b2025-08-20T02:47:49ZengElsevierIndian Pacing and Electrophysiology Journal0972-62922025-01-01251434610.1016/j.ipej.2024.11.001Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism?Arpita Katheria0Kamlesh Raut1Ankit Kumar Sahu2Aditya Kapoor3Department of Cardiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, 226014, Uttar Pradesh, IndiaDepartment of Cardiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, 226014, Uttar Pradesh, IndiaCorresponding author.; Department of Cardiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, 226014, Uttar Pradesh, IndiaDepartment of Cardiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, 226014, Uttar Pradesh, IndiaUnexplained change in parameters, despite good lead positions or odd ECG patterns following implant should lead to suspicion of lead swap, amongst other possibilities. An easy way to confirm this is by paying careful attention to device EGMs and recording the ECG by switching off one channel, and pacing single channel (in either AAI or VVI mode).http://www.sciencedirect.com/science/article/pii/S0972629224001529Dual chamber pacemakerDevice interrogationLead switching |
| spellingShingle | Arpita Katheria Kamlesh Raut Ankit Kumar Sahu Aditya Kapoor Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism? Indian Pacing and Electrophysiology Journal Dual chamber pacemaker Device interrogation Lead switching |
| title | Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism? |
| title_full | Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism? |
| title_fullStr | Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism? |
| title_full_unstemmed | Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism? |
| title_short | Inadvertent higher rate atrial pacing after dual chamber pacemaker implantation - What is the underlying mechanism? |
| title_sort | inadvertent higher rate atrial pacing after dual chamber pacemaker implantation what is the underlying mechanism |
| topic | Dual chamber pacemaker Device interrogation Lead switching |
| url | http://www.sciencedirect.com/science/article/pii/S0972629224001529 |
| work_keys_str_mv | AT arpitakatheria inadvertenthigherrateatrialpacingafterdualchamberpacemakerimplantationwhatistheunderlyingmechanism AT kamleshraut inadvertenthigherrateatrialpacingafterdualchamberpacemakerimplantationwhatistheunderlyingmechanism AT ankitkumarsahu inadvertenthigherrateatrialpacingafterdualchamberpacemakerimplantationwhatistheunderlyingmechanism AT adityakapoor inadvertenthigherrateatrialpacingafterdualchamberpacemakerimplantationwhatistheunderlyingmechanism |