Biotronik Implantable Cardioverter Defibrillator for Measurement of Intrathoracic Impedance: A Report of Two Cases
ABSTRACT Objectives Two patients were received ICDs (Biotronic Iforia7 VR‐TDX). Holter analysis revealed pacing spikes, which occurred regularly at 100 ms intervals following QRS complexes, with consecutive 1024 test pulses at hourly intervals during a specific time, and the VP was 0%. Methods By an...
Saved in:
| Main Authors: | Xuqing Ying, Xinyue Yu, Weixun Cai, Xiaomin Chen, Lihua Wang, Qingcheng Wang, Jie Pang |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Wiley
2024-11-01
|
| Series: | Annals of Noninvasive Electrocardiology |
| Subjects: | |
| Online Access: | https://doi.org/10.1111/anec.70015 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Similar Items
-
Nutritional Status Is Associated with Mortality but Not Appropriate Discharge of Implantable Cardioverter Defibrillators in Patients with Heart Failure
by: Idris Yakut, et al.
Published: (2025-03-01) -
Real-World Efficacy and Safety of the Subcutaneous Implantable Cardioverter Defibrillator: Insights from the GASP Registry
by: Nikias Milaras, et al.
Published: (2025-06-01) -
Dynamic evaluation of atrial fibrillation risk factors using multiple sensors of implantable cardioverter defibrillators
by: Carmelo La Greca, et al.
Published: (2025-08-01) -
Prediction of early mortality after cardioverter-defibrillator implantation for primary prevention of sudden cardiac death: focus on transthoracic echocardiography
by: N. N. Ilov, et al.
Published: (2023-07-01) -
Implantable Cardioverter‐Defibrillator Programming: It Is Time to Get Things Right With a Checklist Manifesto
by: Salvatore Savona, et al.
Published: (2024-08-01)