Design of ultrasonic probes for medical diagnostics

The Mason's equivalent circuit of the piezoelectric transducer is modified in order to investigate separately the influence of parameters of piezoelectric material as well as electrical and acoustical elements applied in the ultrasonic probe. All values transfered to the mechanical side of the...

Full description

Saved in:
Bibliographic Details
Main Authors: G. ŁYPACEWICZ, E. DURIASZ
Format: Article
Language:English
Published: Institute of Fundamental Technological Research Polish Academy of Sciences 2014-05-01
Series:Archives of Acoustics
Online Access:https://acoustics.ippt.pan.pl/index.php/aa/article/view/1296
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849318582268198912
author G. ŁYPACEWICZ
E. DURIASZ
author_facet G. ŁYPACEWICZ
E. DURIASZ
author_sort G. ŁYPACEWICZ
collection DOAJ
description The Mason's equivalent circuit of the piezoelectric transducer is modified in order to investigate separately the influence of parameters of piezoelectric material as well as electrical and acoustical elements applied in the ultrasonic probe. All values transfered to the mechanical side of the electromechanical transformer let us introduce the relative values both electrical and acoustical parameters. The mechanical and dielectric losses of the transducer are taken into consideration. The influence of various parameters of the probe on input admittance and reflected from an ideal reflector pulses is shown.
format Article
id doaj-art-879faede4dc3410c8f03d2cfe07ce121
institution Kabale University
issn 0137-5075
2300-262X
language English
publishDate 2014-05-01
publisher Institute of Fundamental Technological Research Polish Academy of Sciences
record_format Article
series Archives of Acoustics
spelling doaj-art-879faede4dc3410c8f03d2cfe07ce1212025-08-20T03:50:48ZengInstitute of Fundamental Technological Research Polish Academy of SciencesArchives of Acoustics0137-50752300-262X2014-05-01153-4Design of ultrasonic probes for medical diagnosticsG. ŁYPACEWICZ0E. DURIASZ1Ultrasonic Department, Institute of Fundamental Technological Research, Polish Academy of SciencesUltrasonic Department, Institute of Fundamental Technological Research, Polish Academy of SciencesThe Mason's equivalent circuit of the piezoelectric transducer is modified in order to investigate separately the influence of parameters of piezoelectric material as well as electrical and acoustical elements applied in the ultrasonic probe. All values transfered to the mechanical side of the electromechanical transformer let us introduce the relative values both electrical and acoustical parameters. The mechanical and dielectric losses of the transducer are taken into consideration. The influence of various parameters of the probe on input admittance and reflected from an ideal reflector pulses is shown.https://acoustics.ippt.pan.pl/index.php/aa/article/view/1296
spellingShingle G. ŁYPACEWICZ
E. DURIASZ
Design of ultrasonic probes for medical diagnostics
Archives of Acoustics
title Design of ultrasonic probes for medical diagnostics
title_full Design of ultrasonic probes for medical diagnostics
title_fullStr Design of ultrasonic probes for medical diagnostics
title_full_unstemmed Design of ultrasonic probes for medical diagnostics
title_short Design of ultrasonic probes for medical diagnostics
title_sort design of ultrasonic probes for medical diagnostics
url https://acoustics.ippt.pan.pl/index.php/aa/article/view/1296
work_keys_str_mv AT głypacewicz designofultrasonicprobesformedicaldiagnostics
AT eduriasz designofultrasonicprobesformedicaldiagnostics