A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time

Objective: To develop and validate a noninvasive method for measuring tear break-up time (NIBUT) using readily available materials and comparing it to the standard fluorescein tear break-up time (TBUT) assessment. Methods: A prototype xeroscope was made from an ordinary round plastic lid. At its...

Full description

Saved in:
Bibliographic Details
Main Authors: Darien B. Gaw, MD, Bernard Gil O. Tinio, MD
Format: Article
Language:English
Published: Philippine Academy of Ophthalmology 2015-06-01
Series:Philippine Journal of Ophthalmology
Subjects:
Online Access:https://paojournal.com/index.php/pjo/article/view/176
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849338997713666048
author Darien B. Gaw, MD
Bernard Gil O. Tinio, MD
author_facet Darien B. Gaw, MD
Bernard Gil O. Tinio, MD
author_sort Darien B. Gaw, MD
collection DOAJ
description Objective: To develop and validate a noninvasive method for measuring tear break-up time (NIBUT) using readily available materials and comparing it to the standard fluorescein tear break-up time (TBUT) assessment. Methods: A prototype xeroscope was made from an ordinary round plastic lid. At its inner lining, a fluorescent green sticker paper printed with black circular concentric grid pattern was fixed to simulate modified placido rings on the precorneal reflection. A central rectangular cut-out measuring 2 cm x 1 cm was made into the base frame serving as the viewfinder for the device through which the precorneal grid reflection was observed. A 30-LED blue light strip was fixed in the inner rim of the lid using a double-sided adhesive tape and was powered by a multivoltage AC-DC power adapter set at 6-volts at 350mA (2 watts). The prototype device was tested on fifty (50) patients (100 eyes), aged 21 to 65 years old, with no apparent ocular surface disorders and eyelid abnormalities. Measurement of NIBUT was done on both eyes using the prototype xeroscope followed by fluorescein TBUT measurement. The two values were compared and their correlation determined by linear-regression. Results: Fifty (50) patients (100 eyes) were enrolled into the study. The mean age was 48.52 years (range 21 to 62 years). The mean NIBUT for both eyes was 20.10 secs, ranging from 11.56 to 36.17 secs, and the mean TBUT was 18.70 secs, ranging from 10.20 to 30.40 secs. The mean difference between NIBUT and TBUT was 1.40 sec. Regression analysis revealed a strong positive linear correlation between the 2 measurements with a computed R-squared value of 0.9697. Conclusion: The noninvasive technique using the prototype device provided a good alternative approach for measuring tear film stability without disturbing normal tear physiology and dynamics at a fraction of the cost of current commercially-available xeroscopes and tearscopes.
format Article
id doaj-art-850f1788d7ed49e09f3f0a544fc538d2
institution Kabale University
issn 0031-7659
language English
publishDate 2015-06-01
publisher Philippine Academy of Ophthalmology
record_format Article
series Philippine Journal of Ophthalmology
spelling doaj-art-850f1788d7ed49e09f3f0a544fc538d22025-08-20T03:44:14ZengPhilippine Academy of OphthalmologyPhilippine Journal of Ophthalmology0031-76592015-06-014011823176A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up TimeDarien B. Gaw, MD0Bernard Gil O. Tinio, MD1Department of Ophthalmology Makati Medical Center No. 2 Amorsolo Street, Legaspi Village Makati City, PhilippinesDepartment of Ophthalmology Makati Medical Center No. 2 Amorsolo Street, Legaspi Village Makati City, PhilippinesObjective: To develop and validate a noninvasive method for measuring tear break-up time (NIBUT) using readily available materials and comparing it to the standard fluorescein tear break-up time (TBUT) assessment. Methods: A prototype xeroscope was made from an ordinary round plastic lid. At its inner lining, a fluorescent green sticker paper printed with black circular concentric grid pattern was fixed to simulate modified placido rings on the precorneal reflection. A central rectangular cut-out measuring 2 cm x 1 cm was made into the base frame serving as the viewfinder for the device through which the precorneal grid reflection was observed. A 30-LED blue light strip was fixed in the inner rim of the lid using a double-sided adhesive tape and was powered by a multivoltage AC-DC power adapter set at 6-volts at 350mA (2 watts). The prototype device was tested on fifty (50) patients (100 eyes), aged 21 to 65 years old, with no apparent ocular surface disorders and eyelid abnormalities. Measurement of NIBUT was done on both eyes using the prototype xeroscope followed by fluorescein TBUT measurement. The two values were compared and their correlation determined by linear-regression. Results: Fifty (50) patients (100 eyes) were enrolled into the study. The mean age was 48.52 years (range 21 to 62 years). The mean NIBUT for both eyes was 20.10 secs, ranging from 11.56 to 36.17 secs, and the mean TBUT was 18.70 secs, ranging from 10.20 to 30.40 secs. The mean difference between NIBUT and TBUT was 1.40 sec. Regression analysis revealed a strong positive linear correlation between the 2 measurements with a computed R-squared value of 0.9697. Conclusion: The noninvasive technique using the prototype device provided a good alternative approach for measuring tear film stability without disturbing normal tear physiology and dynamics at a fraction of the cost of current commercially-available xeroscopes and tearscopes.https://paojournal.com/index.php/pjo/article/view/176tear break-up timenoninvasive tear break-up timetear film, xeroscopetearscopedry eye syndrome
spellingShingle Darien B. Gaw, MD
Bernard Gil O. Tinio, MD
A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time
Philippine Journal of Ophthalmology
tear break-up time
noninvasive tear break-up time
tear film, xeroscope
tearscope
dry eye syndrome
title A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time
title_full A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time
title_fullStr A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time
title_full_unstemmed A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time
title_short A Simplified Xeroscope for the Noninvasive Measurement of Tear Break-up Time
title_sort simplified xeroscope for the noninvasive measurement of tear break up time
topic tear break-up time
noninvasive tear break-up time
tear film, xeroscope
tearscope
dry eye syndrome
url https://paojournal.com/index.php/pjo/article/view/176
work_keys_str_mv AT darienbgawmd asimplifiedxeroscopeforthenoninvasivemeasurementoftearbreakuptime
AT bernardgilotiniomd asimplifiedxeroscopeforthenoninvasivemeasurementoftearbreakuptime
AT darienbgawmd simplifiedxeroscopeforthenoninvasivemeasurementoftearbreakuptime
AT bernardgilotiniomd simplifiedxeroscopeforthenoninvasivemeasurementoftearbreakuptime