Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula

Purpose. To compare the preoperative measurements of the anterior chamber depth (ACD) by the IOLMaster and Catalys; additionally, to compare the accuracy of the IOL power calculated by the Barrett Universal II formula using the two different measurements. Setting. University of California, Irvine, G...

Full description

Saved in:
Bibliographic Details
Main Authors: Hannah Muniz Castro, Audrey X. Tai, Samuel J. Sampson, Matthew Wade, Marjan Farid, Sumit Garg
Format: Article
Language:English
Published: Wiley 2019-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2019/8172615
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850213570065727488
author Hannah Muniz Castro
Audrey X. Tai
Samuel J. Sampson
Matthew Wade
Marjan Farid
Sumit Garg
author_facet Hannah Muniz Castro
Audrey X. Tai
Samuel J. Sampson
Matthew Wade
Marjan Farid
Sumit Garg
author_sort Hannah Muniz Castro
collection DOAJ
description Purpose. To compare the preoperative measurements of the anterior chamber depth (ACD) by the IOLMaster and Catalys; additionally, to compare the accuracy of the IOL power calculated by the Barrett Universal II formula using the two different measurements. Setting. University of California, Irvine, Gavin Herbert Eye Institute in Irvine, California. Design. Retrospective comparative study. Methods. This study included 144 eyes of 90 patients with a mean age of 72.0 years (range 40.8 to 92.1 years) that underwent femtosecond laser-assisted cataract surgery using Catalys. Preoperative measurements of ACD were taken by the IOLMaster and Catalys. Manifest refraction and refractive spherical equivalent were measured 1 month postoperatively. Expected refractive results were compared with actual postoperative refractive results. Results. The correlation between the ACD values from the two devices was good (r = 0.80). The Catalys ACD measurements yielded a larger ACD compared to the IOLMaster, with a mean difference of 0.22 mm (P<0.0001). The correlation between the postoperative and predicted RSE of the implanted IOL power was excellent (r = 0.96). There was no statistically significant difference between the mean absolute error derived from the IOLMaster, 0.37 diopter (D) ± 0.34 (SD), and the Catalys, 0.37 ± 0.35 D (P=0.50). Conclusions. The Catalys biometry yielded a significantly larger ACD value than the IOLMaster. This difference in ACD value, however, did not reflect in a statistically significant difference in IOL power calculation and refractive prediction error using the Barrett Universal II Formula.
format Article
id doaj-art-83a026ccdf844379b4a1e33a6c9f9dc3
institution OA Journals
issn 2090-004X
2090-0058
language English
publishDate 2019-01-01
publisher Wiley
record_format Article
series Journal of Ophthalmology
spelling doaj-art-83a026ccdf844379b4a1e33a6c9f9dc32025-08-20T02:09:07ZengWileyJournal of Ophthalmology2090-004X2090-00582019-01-01201910.1155/2019/81726158172615Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II FormulaHannah Muniz Castro0Audrey X. Tai1Samuel J. Sampson2Matthew Wade3Marjan Farid4Sumit Garg5UT Health Science Center, Houston, TX, USAUC Irvine Department of Ophthalmology, Gavin Herbert Eye Institute, Irvine, CA, USAKaiser Permanente Los Angeles Medical Center, Los Angeles, CA, USAUC Irvine Department of Ophthalmology, Gavin Herbert Eye Institute, Irvine, CA, USAUC Irvine Department of Ophthalmology, Gavin Herbert Eye Institute, Irvine, CA, USAUC Irvine Department of Ophthalmology, Gavin Herbert Eye Institute, Irvine, CA, USAPurpose. To compare the preoperative measurements of the anterior chamber depth (ACD) by the IOLMaster and Catalys; additionally, to compare the accuracy of the IOL power calculated by the Barrett Universal II formula using the two different measurements. Setting. University of California, Irvine, Gavin Herbert Eye Institute in Irvine, California. Design. Retrospective comparative study. Methods. This study included 144 eyes of 90 patients with a mean age of 72.0 years (range 40.8 to 92.1 years) that underwent femtosecond laser-assisted cataract surgery using Catalys. Preoperative measurements of ACD were taken by the IOLMaster and Catalys. Manifest refraction and refractive spherical equivalent were measured 1 month postoperatively. Expected refractive results were compared with actual postoperative refractive results. Results. The correlation between the ACD values from the two devices was good (r = 0.80). The Catalys ACD measurements yielded a larger ACD compared to the IOLMaster, with a mean difference of 0.22 mm (P<0.0001). The correlation between the postoperative and predicted RSE of the implanted IOL power was excellent (r = 0.96). There was no statistically significant difference between the mean absolute error derived from the IOLMaster, 0.37 diopter (D) ± 0.34 (SD), and the Catalys, 0.37 ± 0.35 D (P=0.50). Conclusions. The Catalys biometry yielded a significantly larger ACD value than the IOLMaster. This difference in ACD value, however, did not reflect in a statistically significant difference in IOL power calculation and refractive prediction error using the Barrett Universal II Formula.http://dx.doi.org/10.1155/2019/8172615
spellingShingle Hannah Muniz Castro
Audrey X. Tai
Samuel J. Sampson
Matthew Wade
Marjan Farid
Sumit Garg
Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula
Journal of Ophthalmology
title Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula
title_full Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula
title_fullStr Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula
title_full_unstemmed Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula
title_short Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula
title_sort accuracy of intraocular lens power calculation using anterior chamber depth from two devices with barrett universal ii formula
url http://dx.doi.org/10.1155/2019/8172615
work_keys_str_mv AT hannahmunizcastro accuracyofintraocularlenspowercalculationusinganteriorchamberdepthfromtwodeviceswithbarrettuniversaliiformula
AT audreyxtai accuracyofintraocularlenspowercalculationusinganteriorchamberdepthfromtwodeviceswithbarrettuniversaliiformula
AT samueljsampson accuracyofintraocularlenspowercalculationusinganteriorchamberdepthfromtwodeviceswithbarrettuniversaliiformula
AT matthewwade accuracyofintraocularlenspowercalculationusinganteriorchamberdepthfromtwodeviceswithbarrettuniversaliiformula
AT marjanfarid accuracyofintraocularlenspowercalculationusinganteriorchamberdepthfromtwodeviceswithbarrettuniversaliiformula
AT sumitgarg accuracyofintraocularlenspowercalculationusinganteriorchamberdepthfromtwodeviceswithbarrettuniversaliiformula