Touchless Mechanism to Detect Rhodamine B Concentration in Distilled Water Using Fiber Bundle
Detection of rhodamine B concentration in distilled water can be conducted without direct contact between the sensor probe and the sample using fiber bundle as a sensor probe and concave mirror as a sample container. The fiber bundle used is a concentric and a pair bundle probe. The working mechanis...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2019-01-01
|
Series: | International Journal of Optics |
Online Access: | http://dx.doi.org/10.1155/2019/5918958 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Detection of rhodamine B concentration in distilled water can be conducted without direct contact between the sensor probe and the sample using fiber bundle as a sensor probe and concave mirror as a sample container. The fiber bundle used is a concentric and a pair bundle probe. The working mechanism of the sensor is based on the displacement sensor and absorption of rhodamine B on laser light at a wavelength of 543 nm. The concentration of rhodamine B was detected through the valley and the peak voltage resulting from the displacement of the sensor probe to the sample surface. The best performance is shown by a sensor that uses a concentric bundle as a probe with a detection range of 0–20 ppm with a resolution of 0.6 ppm. |
---|---|
ISSN: | 1687-9384 1687-9392 |