Evaluation of point-of-care tests for detecting microalbuminuria in diabetic patients

Background: Microalbuminuria, the presence of low levels of albumin in the urine, indicates renal damage and is recognised as a risk factor for the progression of renal and cardiovascular disease. Several international scientific bodies recommend microalbuminuria screening. Point-of-care testing (PO...

Full description

Saved in:
Bibliographic Details
Main Authors: M.M. Lloyd, J. Kuyl, H. van Jaarsveld
Format: Article
Language:English
Published: AOSIS 2011-06-01
Series:South African Family Practice
Subjects:
Online Access:https://safpj.co.za/index.php/safpj/article/view/1632
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background: Microalbuminuria, the presence of low levels of albumin in the urine, indicates renal damage and is recognised as a risk factor for the progression of renal and cardiovascular disease. Several international scientific bodies recommend microalbuminuria screening. Point-of-care testing (POCT) of microalbuminuria allows immediate identification of risk, and monitoring of treatment effects. In this study, two POCT instruments were evaluated as microalbuminuria screening methods. Method: Spot urine specimens from diabetic patients were analysed with the quantitative HemoCue® urine albumin analyser (n = 245), and the semiquantitative Clinitek® microalbumin urine dipstick (n = 204). These results were compared to the respective data for laboratory-determined albumin (nephelometry), creatinine (modified Jaffe) and albumin-to-creatinine ratio (ACR). Results: Linear regression analysis demonstrated a good correlation for the HemoCue® urine albumin with the laboratorydetermined albumin concentration (y = 0.8557x + 0.2487y, r = 0.97). The sensitivities for the HemoCue® and Clinitek® POCT systems were 79.6% and 83.8%, and the specificities 97.1% and 93.8% respectively. Positive and negative predictive values for the HemoCue® were 95.6% and 85.8%, and were 88.6% and 91.0% the Clinitek®. The repeatability of both instruments was excellent. Both instruments are easy to use, and more cost-effective than the laboratory methods for albumin and ACR. Conclusion: Both the HemoCue® and the Clinitek® microalbumin POCT systems for albuminuria are easy to use and inexpensive, and are adequately accurate as a screening method. Although the HemoCue® POCT system measures only urine albumin concentration, its sensitivity and specificity compared well with that of the Clinitek® POCT system, which determines the ACR.
ISSN:2078-6190
2078-6204