Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure

<b>Objectives:</b> The purpose of this study was to analyze image quality and radiation exposure of thoracic computed tomography angiography (CTA) in children with congenital heart diseases (CHDs) using either manual contrast medium (CM) injection or automated power injectors with bolus...

Full description

Saved in:
Bibliographic Details
Main Authors: Jochen Pfeifer, Deborah Driulini, Katrin Altmeyer, Gudrun Wagenpfeil, Martin Poryo, Christian Giebels, Arno Bücker, Alexander Massmann, Hashim Abdul-Khaliq, Peter Fries
Format: Article
Language:English
Published: MDPI AG 2025-04-01
Series:Diagnostics
Subjects:
Online Access:https://www.mdpi.com/2075-4418/15/9/1103
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850155896382947328
author Jochen Pfeifer
Deborah Driulini
Katrin Altmeyer
Gudrun Wagenpfeil
Martin Poryo
Christian Giebels
Arno Bücker
Alexander Massmann
Hashim Abdul-Khaliq
Peter Fries
author_facet Jochen Pfeifer
Deborah Driulini
Katrin Altmeyer
Gudrun Wagenpfeil
Martin Poryo
Christian Giebels
Arno Bücker
Alexander Massmann
Hashim Abdul-Khaliq
Peter Fries
author_sort Jochen Pfeifer
collection DOAJ
description <b>Objectives:</b> The purpose of this study was to analyze image quality and radiation exposure of thoracic computed tomography angiography (CTA) in children with congenital heart diseases (CHDs) using either manual contrast medium (CM) injection or automated power injectors with bolus tracking. <b>Methods:</b> A total of 137 thoracic CTAs of 120 consecutive pediatric patients were included in this retrospective study. We analyzed the method of CM administration (power injection with bolus tracking (PI) or manual injection (MI)), injection routes, volumes and flow rates of CM. For the evaluation of objective image quality, attenuation values in the heart chambers and great thoracic vessels were determined by region-of-interest (ROI) analysis and signal-to-noise (SNR) and contrast-to-noise (CNR) ratios calculated thereof. Visual image quality was assessed by two blinded readers (four-point Likert-scale) analyzing the presence of artifacts and the depiction of relevant anatomical structures. Effective radiation doses were calculated with dose length products and specific conversion factors. <b>Results:</b> CM administration was performed using PI in 119/137 CTAs, whereas MI was conducted in 18/137. The smallest size of peripheral venous cannulas was 24 gauge in 36/137 (26.3%) cases. Overall mean CM volume was 17 mL ± 16 mL (mean ± SD). In PI, the mean flow rate of CM was 1.52 ± 0.90 mL/s with a range between 0.5 and 5.0 mL/s. When comparing the overall PI population and an age-, size- and weight-matched PI subpopulation (18 cases) with the MI population, attenuation values in Hounsfield units (HU) and CNR values were significantly higher in the PI groups than in the MI group for each relevant cardiac structure (left ventricle, right ventricle, ascending aorta and pulmonary trunk, <i>p</i> = 0.02–0.001). Overall image quality and depiction of cardiac structures were rated significantly better in CTAs with PI (interquartile ranges: “good” to “excellent” (Likert 3–4)) in PI compared with CTAs acquired with MI (interquartile ranges: “fair” to “good” (2–3)) in MI by both readers (<i>p</i> < 0.001). The inter-observer reliability was strong, with a Kendall’s Tau-b correlation coefficient of <i>τ</i> = 0.802 (<i>p</i> < 0.001). The mean effective radiation dose (E) did not differ significantly when comparing the stratified samples (i.e., the matched PI subgroup and the MI group; 0.5 (±0.3) mSv in both, <i>p</i> = 0.76). There were no complications associated with the CM injections for both application approaches. <b>Conclusions:</b> Automated contrast agent applications with power injectors and bolus tracking ensure better image quality in pediatric CTA, even when low volumes and flow rates need to be applied. There is a slight increase in radiation associated with bolus tracking. This approach represents a suitable imaging technique for the work-up of congenital heart disease.
format Article
id doaj-art-79c5e8c9c42e4e7d94dcedc06aee6e84
institution OA Journals
issn 2075-4418
language English
publishDate 2025-04-01
publisher MDPI AG
record_format Article
series Diagnostics
spelling doaj-art-79c5e8c9c42e4e7d94dcedc06aee6e842025-08-20T02:24:45ZengMDPI AGDiagnostics2075-44182025-04-01159110310.3390/diagnostics15091103Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation ExposureJochen Pfeifer0Deborah Driulini1Katrin Altmeyer2Gudrun Wagenpfeil3Martin Poryo4Christian Giebels5Arno Bücker6Alexander Massmann7Hashim Abdul-Khaliq8Peter Fries9Department of Pediatric Cardiology, Saarland University Medical Center, 66421 Homburg, GermanyClinic for Diagnostic and Interventional Radiology, Saarland University Medical Center, 66421 Homburg, GermanyClinic for Diagnostic and Interventional Radiology, Saarland University Medical Center, 66421 Homburg, GermanyInstitute of Medical Biometry, Epidemiology and Medical Informatics, Saarland University Medical Center, 66421 Homburg, GermanyDepartment of Pediatric Cardiology, Saarland University Medical Center, 66421 Homburg, GermanyDepartment of Cardiovascular Surgery, University Heart Center Freiburg-Bad Krozingen, 79189 Bad Krozingen, GermanyClinic for Diagnostic and Interventional Radiology, Saarland University Medical Center, 66421 Homburg, GermanyDepartment of Radiology and Nuclear Imaging, Robert Bosch Hospital, 70376 Stuttgart, GermanyDepartment of Pediatric Cardiology, Saarland University Medical Center, 66421 Homburg, GermanyClinic for Diagnostic and Interventional Radiology, Saarland University Medical Center, 66421 Homburg, Germany<b>Objectives:</b> The purpose of this study was to analyze image quality and radiation exposure of thoracic computed tomography angiography (CTA) in children with congenital heart diseases (CHDs) using either manual contrast medium (CM) injection or automated power injectors with bolus tracking. <b>Methods:</b> A total of 137 thoracic CTAs of 120 consecutive pediatric patients were included in this retrospective study. We analyzed the method of CM administration (power injection with bolus tracking (PI) or manual injection (MI)), injection routes, volumes and flow rates of CM. For the evaluation of objective image quality, attenuation values in the heart chambers and great thoracic vessels were determined by region-of-interest (ROI) analysis and signal-to-noise (SNR) and contrast-to-noise (CNR) ratios calculated thereof. Visual image quality was assessed by two blinded readers (four-point Likert-scale) analyzing the presence of artifacts and the depiction of relevant anatomical structures. Effective radiation doses were calculated with dose length products and specific conversion factors. <b>Results:</b> CM administration was performed using PI in 119/137 CTAs, whereas MI was conducted in 18/137. The smallest size of peripheral venous cannulas was 24 gauge in 36/137 (26.3%) cases. Overall mean CM volume was 17 mL ± 16 mL (mean ± SD). In PI, the mean flow rate of CM was 1.52 ± 0.90 mL/s with a range between 0.5 and 5.0 mL/s. When comparing the overall PI population and an age-, size- and weight-matched PI subpopulation (18 cases) with the MI population, attenuation values in Hounsfield units (HU) and CNR values were significantly higher in the PI groups than in the MI group for each relevant cardiac structure (left ventricle, right ventricle, ascending aorta and pulmonary trunk, <i>p</i> = 0.02–0.001). Overall image quality and depiction of cardiac structures were rated significantly better in CTAs with PI (interquartile ranges: “good” to “excellent” (Likert 3–4)) in PI compared with CTAs acquired with MI (interquartile ranges: “fair” to “good” (2–3)) in MI by both readers (<i>p</i> < 0.001). The inter-observer reliability was strong, with a Kendall’s Tau-b correlation coefficient of <i>τ</i> = 0.802 (<i>p</i> < 0.001). The mean effective radiation dose (E) did not differ significantly when comparing the stratified samples (i.e., the matched PI subgroup and the MI group; 0.5 (±0.3) mSv in both, <i>p</i> = 0.76). There were no complications associated with the CM injections for both application approaches. <b>Conclusions:</b> Automated contrast agent applications with power injectors and bolus tracking ensure better image quality in pediatric CTA, even when low volumes and flow rates need to be applied. There is a slight increase in radiation associated with bolus tracking. This approach represents a suitable imaging technique for the work-up of congenital heart disease.https://www.mdpi.com/2075-4418/15/9/1103computed tomography angiographycontrast mediumimage qualitymanual injectionautomated power injectorbolus tracking
spellingShingle Jochen Pfeifer
Deborah Driulini
Katrin Altmeyer
Gudrun Wagenpfeil
Martin Poryo
Christian Giebels
Arno Bücker
Alexander Massmann
Hashim Abdul-Khaliq
Peter Fries
Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure
Diagnostics
computed tomography angiography
contrast medium
image quality
manual injection
automated power injector
bolus tracking
title Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure
title_full Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure
title_fullStr Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure
title_full_unstemmed Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure
title_short Thoracic CT Angiographies in Children Using Automated Power Injection with Bolus Tracking Versus Manual Contrast Injection: Analysis of Contrast Enhancement, Image Quality and Radiation Exposure
title_sort thoracic ct angiographies in children using automated power injection with bolus tracking versus manual contrast injection analysis of contrast enhancement image quality and radiation exposure
topic computed tomography angiography
contrast medium
image quality
manual injection
automated power injector
bolus tracking
url https://www.mdpi.com/2075-4418/15/9/1103
work_keys_str_mv AT jochenpfeifer thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT deborahdriulini thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT katrinaltmeyer thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT gudrunwagenpfeil thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT martinporyo thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT christiangiebels thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT arnobucker thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT alexandermassmann thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT hashimabdulkhaliq thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure
AT peterfries thoracicctangiographiesinchildrenusingautomatedpowerinjectionwithbolustrackingversusmanualcontrastinjectionanalysisofcontrastenhancementimagequalityandradiationexposure