Contact-free magnetic resonance imaging and spectroscopy with acoustic levitation

Abstract Conventional magnetic resonance measurements often rely on the use of sample containers. This limits the implementation of time-resolved studies at the molecular level of liquid samples undergoing evaporation or other dynamic phenomena that require access to the liquid-gas interface. In thi...

Full description

Saved in:
Bibliographic Details
Main Authors: Smaragda-Maria Argyri, Leo Svenningsson, Feryal Guerroudj, Diana Bernin, Lars Evenäs, Romain Bordes
Format: Article
Language:English
Published: Nature Portfolio 2025-04-01
Series:Nature Communications
Online Access:https://doi.org/10.1038/s41467-025-58949-2
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract Conventional magnetic resonance measurements often rely on the use of sample containers. This limits the implementation of time-resolved studies at the molecular level of liquid samples undergoing evaporation or other dynamic phenomena that require access to the liquid-gas interface. In this study, we developed a demagnetized acoustic levitator to perform magnetic resonance studies on liquid samples, in a contact-free manner. The performance of the levitator inside a 7.05 T magnetic field was examined and magnetic resonance images of the levitator and the levitated samples were acquired. Then, we collected magnetic resonance spectra of the levitated droplets by applying localized and non-localized pulse sequences and we examined the effect of the droplet shape on the chemical shift. Additionally, we conducted time-resolved experiments on pure solvents and mixtures, and captured physical and chemical molecular interactions, in real-time. This approach enables contact-free studies at the molecular level of dynamic phenomena on a microliter droplet using magnetic resonance techniques.
ISSN:2041-1723