Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality
The variation of the bulk etch rate (VB) of nuclear track detector CR-39 with the normality of the chemical etchant were studied. The removal thickness measurement method is used to determine the bulk etch rate. the detectors were etched by an aqueous solution of NaOH with normalities (4, 5, 6, 7)...
Saved in:
| Main Authors: | , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Tikrit University
2023-02-01
|
| Series: | Tikrit Journal of Pure Science |
| Subjects: | |
| Online Access: | https://tjpsj.org/index.php/tjps/article/view/1250 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| _version_ | 1849427553342717952 |
|---|---|
| author | Ahmed A. Al-obedy Ammar A. AL-Battawy Omar K. Younus |
| author_facet | Ahmed A. Al-obedy Ammar A. AL-Battawy Omar K. Younus |
| author_sort | Ahmed A. Al-obedy |
| collection | DOAJ |
| description |
The variation of the bulk etch rate (VB) of nuclear track detector CR-39 with the normality of the chemical etchant were studied. The removal thickness measurement method is used to determine the bulk etch rate. the detectors were etched by an aqueous solution of NaOH with normalities (4, 5, 6, 7) N at temperature (70±1) oC. It was found that the bulk etch rate is exponentially increasing with the normality of the solution and its values are (0.687, 1.421, 0.954, 2.082) µm/h at normalities (4, 5, 6, 7) N respectively. the relation between the bulk etch rate and the etchant normality was an exponential relation and the results showed good agreement with the results of the others used other normalities and another measurement method.
|
| format | Article |
| id | doaj-art-7006a9dc851f4ff9902f2e2e793160eb |
| institution | Kabale University |
| issn | 1813-1662 2415-1726 |
| language | English |
| publishDate | 2023-02-01 |
| publisher | Tikrit University |
| record_format | Article |
| series | Tikrit Journal of Pure Science |
| spelling | doaj-art-7006a9dc851f4ff9902f2e2e793160eb2025-08-20T03:28:59ZengTikrit UniversityTikrit Journal of Pure Science1813-16622415-17262023-02-0120510.25130/tjps.v20i5.1250Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution NormalityAhmed A. Al-obedyAmmar A. AL-BattawyOmar K. Younus The variation of the bulk etch rate (VB) of nuclear track detector CR-39 with the normality of the chemical etchant were studied. The removal thickness measurement method is used to determine the bulk etch rate. the detectors were etched by an aqueous solution of NaOH with normalities (4, 5, 6, 7) N at temperature (70±1) oC. It was found that the bulk etch rate is exponentially increasing with the normality of the solution and its values are (0.687, 1.421, 0.954, 2.082) µm/h at normalities (4, 5, 6, 7) N respectively. the relation between the bulk etch rate and the etchant normality was an exponential relation and the results showed good agreement with the results of the others used other normalities and another measurement method. https://tjpsj.org/index.php/tjps/article/view/1250VariationBulk Etch RateNuclear Track DetectorCR-39Chemical Etchant Solution |
| spellingShingle | Ahmed A. Al-obedy Ammar A. AL-Battawy Omar K. Younus Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality Tikrit Journal of Pure Science Variation Bulk Etch Rate Nuclear Track Detector CR-39 Chemical Etchant Solution |
| title | Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality |
| title_full | Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality |
| title_fullStr | Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality |
| title_full_unstemmed | Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality |
| title_short | Variation of the Bulk Etch Rate of Nuclear Track Detector CR-39 With the Chemical Etchant Solution Normality |
| title_sort | variation of the bulk etch rate of nuclear track detector cr 39 with the chemical etchant solution normality |
| topic | Variation Bulk Etch Rate Nuclear Track Detector CR-39 Chemical Etchant Solution |
| url | https://tjpsj.org/index.php/tjps/article/view/1250 |
| work_keys_str_mv | AT ahmedaalobedy variationofthebulketchrateofnucleartrackdetectorcr39withthechemicaletchantsolutionnormality AT ammaraalbattawy variationofthebulketchrateofnucleartrackdetectorcr39withthechemicaletchantsolutionnormality AT omarkyounus variationofthebulketchrateofnucleartrackdetectorcr39withthechemicaletchantsolutionnormality |