Effect of fluid contamination on the removal torque of abutment screws

Abstract Background Screw loosening is one of the most common mechanical complications of implant-supported restorations. The purpose of this study was to evaluate the effects of contamination with different liquids on the removal torque values (RTVs) in the implant‒abutment connection. Methods A to...

Full description

Saved in:
Bibliographic Details
Main Authors: Tuba Toprak, Göknil Alkan Demetoğlu
Format: Article
Language:English
Published: BMC 2025-05-01
Series:BMC Oral Health
Subjects:
Online Access:https://doi.org/10.1186/s12903-025-06130-y
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849309458028560384
author Tuba Toprak
Göknil Alkan Demetoğlu
author_facet Tuba Toprak
Göknil Alkan Demetoğlu
author_sort Tuba Toprak
collection DOAJ
description Abstract Background Screw loosening is one of the most common mechanical complications of implant-supported restorations. The purpose of this study was to evaluate the effects of contamination with different liquids on the removal torque values (RTVs) in the implant‒abutment connection. Methods A total of 66 DTI implants (6 groups, 11 implants in each group) were used. The screw holes of 55 implants were contaminated with either blood from a 29-year-old female donor (BG), saliva (SG), a blood and saliva mixture (BS), blood and Ankaferd Blood Stopper mixture (BA), topical anesthetic (TA), and the remaining noncontaminated screws served as a control group (NC). Afterwards, the abutment screws were tightened to the implants via a digital torque meter (DID-4 A) with a torque of 20 Ncm. RTVs were recorded after thermocycling. The data obtained from the study were statistically evaluated via the Shapiro‒Wilk test, the Levene test, one-way analysis of variance and the Bonferroni correction. Results When the average removal torque of the groups were examined, the group with the highest average value was the control group. All contaminating agents in clinical conditions caused screw loosening. Conclusion Clinicians should consider that RTVs ​​may be affected by contamination of the implant abutment screw hole with blood, saliva, hemostatic agent or local anesthetic solution.
format Article
id doaj-art-6dfbc09ed87d495087720096a8b8481f
institution Kabale University
issn 1472-6831
language English
publishDate 2025-05-01
publisher BMC
record_format Article
series BMC Oral Health
spelling doaj-art-6dfbc09ed87d495087720096a8b8481f2025-08-20T03:54:10ZengBMCBMC Oral Health1472-68312025-05-012511610.1186/s12903-025-06130-yEffect of fluid contamination on the removal torque of abutment screwsTuba Toprak0Göknil Alkan Demetoğlu1Faculty of Dentistry, Department of Prosthetic Dentistry, Adnan Menderes UniversityFaculty of Dentistry, Department of Prosthetic Dentistry, Adnan Menderes UniversityAbstract Background Screw loosening is one of the most common mechanical complications of implant-supported restorations. The purpose of this study was to evaluate the effects of contamination with different liquids on the removal torque values (RTVs) in the implant‒abutment connection. Methods A total of 66 DTI implants (6 groups, 11 implants in each group) were used. The screw holes of 55 implants were contaminated with either blood from a 29-year-old female donor (BG), saliva (SG), a blood and saliva mixture (BS), blood and Ankaferd Blood Stopper mixture (BA), topical anesthetic (TA), and the remaining noncontaminated screws served as a control group (NC). Afterwards, the abutment screws were tightened to the implants via a digital torque meter (DID-4 A) with a torque of 20 Ncm. RTVs were recorded after thermocycling. The data obtained from the study were statistically evaluated via the Shapiro‒Wilk test, the Levene test, one-way analysis of variance and the Bonferroni correction. Results When the average removal torque of the groups were examined, the group with the highest average value was the control group. All contaminating agents in clinical conditions caused screw loosening. Conclusion Clinicians should consider that RTVs ​​may be affected by contamination of the implant abutment screw hole with blood, saliva, hemostatic agent or local anesthetic solution.https://doi.org/10.1186/s12903-025-06130-yAbutmentHemostatic agentRemoval torque valueScrew loosening
spellingShingle Tuba Toprak
Göknil Alkan Demetoğlu
Effect of fluid contamination on the removal torque of abutment screws
BMC Oral Health
Abutment
Hemostatic agent
Removal torque value
Screw loosening
title Effect of fluid contamination on the removal torque of abutment screws
title_full Effect of fluid contamination on the removal torque of abutment screws
title_fullStr Effect of fluid contamination on the removal torque of abutment screws
title_full_unstemmed Effect of fluid contamination on the removal torque of abutment screws
title_short Effect of fluid contamination on the removal torque of abutment screws
title_sort effect of fluid contamination on the removal torque of abutment screws
topic Abutment
Hemostatic agent
Removal torque value
Screw loosening
url https://doi.org/10.1186/s12903-025-06130-y
work_keys_str_mv AT tubatoprak effectoffluidcontaminationontheremovaltorqueofabutmentscrews
AT goknilalkandemetoglu effectoffluidcontaminationontheremovaltorqueofabutmentscrews