Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis
Purpose. Comparison of monoaxial and polyaxial screws with the use of subcutaneous anterior pelvic fixation. Methods. Four different groups each having 5 constructs were tested in distraction within the elastic range. Once that was completed, 3 components were tested in torsion within the elastic ra...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2013-01-01
|
Series: | Advances in Orthopedics |
Online Access: | http://dx.doi.org/10.1155/2013/683120 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832563781162500096 |
---|---|
author | Rahul Vaidya Ndidi Onwudiwe Matthew Roth Anil Sethi |
author_facet | Rahul Vaidya Ndidi Onwudiwe Matthew Roth Anil Sethi |
author_sort | Rahul Vaidya |
collection | DOAJ |
description | Purpose. Comparison of monoaxial and polyaxial screws with the use of subcutaneous anterior pelvic fixation. Methods. Four different groups each having 5 constructs were tested in distraction within the elastic range. Once that was completed, 3 components were tested in torsion within the elastic range, 2 to torsional failure and 3 in distraction until failure. Results. The pedicle screw systems showed higher stiffness (4.008 ± 0.113 Nmm monoaxial, 3.638 ± 0.108 Nmm Click-x; 3.634 ± 0.147 Nmm Pangea) than the exfix system (2.882 ± 0.054 Nmm) in distraction. In failure testing, monoaxial pedicle screw system was stronger (360 N) than exfixes (160 N) and polyaxial devices which failed if distracted greater than 4 cm (157 N Click-x or 138 N Pangea). The exfix had higher peak torque and torsional stiffness than all pedicle systems. In torsion, the yield strengths were the same for all constructs. Conclusion. The infix device constructed with polyaxial or monoaxial pedicle screws is stiffer than the 2 pin external fixator in distraction testing. In extreme cases, the use of reinforcement or monoaxial systems which do not fail even at 360 N is a better option. In torsional testing, the 2 pin external fixator is stiffer than the pedicle screw systems. |
format | Article |
id | doaj-art-6df175390808465188f09ff8bd613167 |
institution | Kabale University |
issn | 2090-3464 2090-3472 |
language | English |
publishDate | 2013-01-01 |
publisher | Wiley |
record_format | Article |
series | Advances in Orthopedics |
spelling | doaj-art-6df175390808465188f09ff8bd6131672025-02-03T01:12:31ZengWileyAdvances in Orthopedics2090-34642090-34722013-01-01201310.1155/2013/683120683120Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical AnalysisRahul Vaidya0Ndidi Onwudiwe1Matthew Roth2Anil Sethi3Detroit Medical Center, Orthopedics Department, Wayne State School of Medicine, 4D UHC, St. Antoine Street, 540 E Canfield Street, Detroit, MI 48201, USADetroit Medical Center, Orthopedics Department, Wayne State School of Medicine, 4D UHC, St. Antoine Street, 540 E Canfield Street, Detroit, MI 48201, USADetroit Medical Center, Orthopedics Department, Wayne State School of Medicine, 4D UHC, St. Antoine Street, 540 E Canfield Street, Detroit, MI 48201, USADetroit Medical Center, Orthopedics Department, Wayne State School of Medicine, 4D UHC, St. Antoine Street, 540 E Canfield Street, Detroit, MI 48201, USAPurpose. Comparison of monoaxial and polyaxial screws with the use of subcutaneous anterior pelvic fixation. Methods. Four different groups each having 5 constructs were tested in distraction within the elastic range. Once that was completed, 3 components were tested in torsion within the elastic range, 2 to torsional failure and 3 in distraction until failure. Results. The pedicle screw systems showed higher stiffness (4.008 ± 0.113 Nmm monoaxial, 3.638 ± 0.108 Nmm Click-x; 3.634 ± 0.147 Nmm Pangea) than the exfix system (2.882 ± 0.054 Nmm) in distraction. In failure testing, monoaxial pedicle screw system was stronger (360 N) than exfixes (160 N) and polyaxial devices which failed if distracted greater than 4 cm (157 N Click-x or 138 N Pangea). The exfix had higher peak torque and torsional stiffness than all pedicle systems. In torsion, the yield strengths were the same for all constructs. Conclusion. The infix device constructed with polyaxial or monoaxial pedicle screws is stiffer than the 2 pin external fixator in distraction testing. In extreme cases, the use of reinforcement or monoaxial systems which do not fail even at 360 N is a better option. In torsional testing, the 2 pin external fixator is stiffer than the pedicle screw systems.http://dx.doi.org/10.1155/2013/683120 |
spellingShingle | Rahul Vaidya Ndidi Onwudiwe Matthew Roth Anil Sethi Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis Advances in Orthopedics |
title | Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis |
title_full | Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis |
title_fullStr | Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis |
title_full_unstemmed | Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis |
title_short | Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis |
title_sort | monoaxial pedicle screws are superior to polyaxial pedicle screws and the two pin external fixator for subcutaneous anterior pelvic fixation in a biomechanical analysis |
url | http://dx.doi.org/10.1155/2013/683120 |
work_keys_str_mv | AT rahulvaidya monoaxialpediclescrewsaresuperiortopolyaxialpediclescrewsandthetwopinexternalfixatorforsubcutaneousanteriorpelvicfixationinabiomechanicalanalysis AT ndidionwudiwe monoaxialpediclescrewsaresuperiortopolyaxialpediclescrewsandthetwopinexternalfixatorforsubcutaneousanteriorpelvicfixationinabiomechanicalanalysis AT matthewroth monoaxialpediclescrewsaresuperiortopolyaxialpediclescrewsandthetwopinexternalfixatorforsubcutaneousanteriorpelvicfixationinabiomechanicalanalysis AT anilsethi monoaxialpediclescrewsaresuperiortopolyaxialpediclescrewsandthetwopinexternalfixatorforsubcutaneousanteriorpelvicfixationinabiomechanicalanalysis |