THE INVESTIGATION OF THE INTERACTION OF THE IMPLANTABLE ROTARY BLOOD PUMP AND THE LEFT VENTRICLE ON THE MOCK CIRCULATION SYSTEM

We analyzed the dynamics of the interaction of rotary pump, connected to the scheme «the left ventricle – aor- ta», and left ventricle in norms and heart failure, using data obtained on the mock circulatory system. Adverse pumping states such as suction and regurgitation can be detected. These limit...

Full description

Saved in:
Bibliographic Details
Main Authors: E. G. Konysheva, G. P. Itkin, K. N. Dozorov, V. L. Kudinov, D. V. Shumakov
Format: Article
Language:Russian
Published: Federal Research Center of Transplantology and Artificial Organs named after V.I.Shumakov 2010-06-01
Series:Вестник трансплантологии и искусственных органов
Subjects:
Online Access:https://journal.transpl.ru/vtio/article/view/332
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We analyzed the dynamics of the interaction of rotary pump, connected to the scheme «the left ventricle – aor- ta», and left ventricle in norms and heart failure, using data obtained on the mock circulatory system. Adverse pumping states such as suction and regurgitation can be detected. These limits depend on the degree of cardiac failure, which confirms the need to control the speed of rotation of the impeller in the range of the cardiac failure from pathology to normal. It has been shown that the pulsating pressure at the inlet of the pump modulates the flow of the pump, making it a pulsating, what positively affects on the dynamics of cardiac output.
ISSN:1995-1191