3D-Printed Anthropomorphic Phantom Framework for Spectral Photon-Counting CT: A Breast Phantom Application

Anthropomorphic phantoms are critical in computed tomography (CT) imaging, simulating human anatomy for image quality and dosimetry evaluation. Traditional phantoms, however, often fail to accurately replicate complex tissue patterns, particularly for advanced technologies like spectral photon-count...

Full description

Saved in:
Bibliographic Details
Main Authors: Kyriaki Katsikari, Nadine Francis, Nabil Maalej, Charalampos Pitsalidis, Aamir Younis Raja
Format: Article
Language:English
Published: IEEE 2025-01-01
Series:IEEE Access
Subjects:
Online Access:https://ieeexplore.ieee.org/document/10812723/
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Anthropomorphic phantoms are critical in computed tomography (CT) imaging, simulating human anatomy for image quality and dosimetry evaluation. Traditional phantoms, however, often fail to accurately replicate complex tissue patterns, particularly for advanced technologies like spectral photon-counting CT (SPCCT). This study introduces a framework for creating 3D-printed anthropomorphic phantoms optimized for SPCCT, beginning with a breast phantom. A grid search optimization method was used to generate an optimal image acquisition protocol, capable of distinguishing between adipose, fibroglandular, skin, and carcinoma tissues. Using the MARS Microlab <inline-formula> <tex-math notation="LaTeX">$5\times 120$ </tex-math></inline-formula> SPCCT system, it was found that a tube voltage of 80 kV and narrow energy bins 7-27 keV, 27-32 keV, 32-37 keV, 37-42 keV, and 42-80 keV, offered superior results. Various photopolymerizable resins and thermoplastic filaments were tested as tissue-mimicking materials (TMMs), comparing their linear attenuation coefficients with those of the tissues and reference TMMs. The breast phantom was created by segmenting an MRI image from the Duke Breast Cancer MRI dataset into two models: one for skin, fibroglandular, and carcinoma tissues, 3D-printed with e-PLA resin, and another for adipose tissue, replicated using paraffin wax. Multi-energy images of the phantom were then evaluated using both narrow and wide energy bins for validation. While this study focused on a breast phantom, the framework can be applied to create various anthropomorphic phantoms, enhancing SPCCT imaging and potentially improving diagnostic accuracy and patient outcomes.
ISSN:2169-3536