Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production

Reduced sulfur compounds are considered to be important odorants from pig production due to their low odor threshold values and low solubility in slurry. The objective of the present study was to investigate the use of a portable method with a single silica gel column for trapping/separation coupled...

Full description

Saved in:
Bibliographic Details
Main Authors: Michael Jørgen Hansen, Kei Toda, Tomoaki Obata, Anders Peter S. Adamsen, Anders Feilberg
Format: Article
Language:English
Published: Wiley 2012-01-01
Series:Journal of Analytical Methods in Chemistry
Online Access:http://dx.doi.org/10.1155/2012/489239
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849411936060440576
author Michael Jørgen Hansen
Kei Toda
Tomoaki Obata
Anders Peter S. Adamsen
Anders Feilberg
author_facet Michael Jørgen Hansen
Kei Toda
Tomoaki Obata
Anders Peter S. Adamsen
Anders Feilberg
author_sort Michael Jørgen Hansen
collection DOAJ
description Reduced sulfur compounds are considered to be important odorants from pig production due to their low odor threshold values and low solubility in slurry. The objective of the present study was to investigate the use of a portable method with a single silica gel column for trapping/separation coupled with chemiluminescence detection (SCTS-CL) for measurement of methanethiol and dimethyl sulfide in sample air from pig production. Proton-transfer-reaction mass spectrometry (PTR-MS) was used to evaluate the trapping/separation. The silica gel column used for the SCTS-CL efficiently collected hydrogen sulfide, methanethiol and dimethyl sulfide. The measurement of methanethiol by SCTS-CL was clearly interfered by the high concentration of hydrogen sulfide found in pig production, and a removal of hydrogen sulfide was necessary to obtain reliable results. Air samples taken from a facility with growing-finishing pigs were analyzed by SCTS-CL, PTR-MS, and a gas chromatograph with sulfur chemiluminescence detection (GC-SCD) to evaluate the SCTS-CL. The difference between the concentrations of methanethiol and dimethyl sulfide measured with SCTS-CL, PTR-MS, and GC-SCD was below 10%. In conclusion, the SCTS-CL is a portable and low-cost alternative to the commercial methods that can be used to measure methanethiol and dimethyl sulfide in sample air from pig production.
format Article
id doaj-art-6554b33d3bd443f6980dbf1a58816ec8
institution Kabale University
issn 2090-8865
2090-8873
language English
publishDate 2012-01-01
publisher Wiley
record_format Article
series Journal of Analytical Methods in Chemistry
spelling doaj-art-6554b33d3bd443f6980dbf1a58816ec82025-08-20T03:34:36ZengWileyJournal of Analytical Methods in Chemistry2090-88652090-88732012-01-01201210.1155/2012/489239489239Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig ProductionMichael Jørgen Hansen0Kei Toda1Tomoaki Obata2Anders Peter S. Adamsen3Anders Feilberg4Department of Engineering, Science and Technology, Aarhus University, Blichers Allé 20, 8830 Tjele, DenmarkDepartment of Chemistry, Kumamoto University, 2-39-1 Kurokami, Kumamoto 860-8555, JapanDepartment of Chemistry, Kumamoto University, 2-39-1 Kurokami, Kumamoto 860-8555, JapanDepartment of Engineering, Science and Technology, Aarhus University, Blichers Allé 20, 8830 Tjele, DenmarkDepartment of Engineering, Science and Technology, Aarhus University, Blichers Allé 20, 8830 Tjele, DenmarkReduced sulfur compounds are considered to be important odorants from pig production due to their low odor threshold values and low solubility in slurry. The objective of the present study was to investigate the use of a portable method with a single silica gel column for trapping/separation coupled with chemiluminescence detection (SCTS-CL) for measurement of methanethiol and dimethyl sulfide in sample air from pig production. Proton-transfer-reaction mass spectrometry (PTR-MS) was used to evaluate the trapping/separation. The silica gel column used for the SCTS-CL efficiently collected hydrogen sulfide, methanethiol and dimethyl sulfide. The measurement of methanethiol by SCTS-CL was clearly interfered by the high concentration of hydrogen sulfide found in pig production, and a removal of hydrogen sulfide was necessary to obtain reliable results. Air samples taken from a facility with growing-finishing pigs were analyzed by SCTS-CL, PTR-MS, and a gas chromatograph with sulfur chemiluminescence detection (GC-SCD) to evaluate the SCTS-CL. The difference between the concentrations of methanethiol and dimethyl sulfide measured with SCTS-CL, PTR-MS, and GC-SCD was below 10%. In conclusion, the SCTS-CL is a portable and low-cost alternative to the commercial methods that can be used to measure methanethiol and dimethyl sulfide in sample air from pig production.http://dx.doi.org/10.1155/2012/489239
spellingShingle Michael Jørgen Hansen
Kei Toda
Tomoaki Obata
Anders Peter S. Adamsen
Anders Feilberg
Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production
Journal of Analytical Methods in Chemistry
title Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production
title_full Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production
title_fullStr Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production
title_full_unstemmed Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production
title_short Evaluation of Single Column Trapping/Separation and Chemiluminescence Detection for Measurement of Methanethiol and Dimethyl Sulfide from Pig Production
title_sort evaluation of single column trapping separation and chemiluminescence detection for measurement of methanethiol and dimethyl sulfide from pig production
url http://dx.doi.org/10.1155/2012/489239
work_keys_str_mv AT michaeljørgenhansen evaluationofsinglecolumntrappingseparationandchemiluminescencedetectionformeasurementofmethanethiolanddimethylsulfidefrompigproduction
AT keitoda evaluationofsinglecolumntrappingseparationandchemiluminescencedetectionformeasurementofmethanethiolanddimethylsulfidefrompigproduction
AT tomoakiobata evaluationofsinglecolumntrappingseparationandchemiluminescencedetectionformeasurementofmethanethiolanddimethylsulfidefrompigproduction
AT anderspetersadamsen evaluationofsinglecolumntrappingseparationandchemiluminescencedetectionformeasurementofmethanethiolanddimethylsulfidefrompigproduction
AT andersfeilberg evaluationofsinglecolumntrappingseparationandchemiluminescencedetectionformeasurementofmethanethiolanddimethylsulfidefrompigproduction