Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure
This paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered ful...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2014-01-01
|
Series: | Case Reports in Orthopedics |
Online Access: | http://dx.doi.org/10.1155/2014/896348 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832561845950480384 |
---|---|
author | Girish N. Swamy Conal Quah Elmunzar Bagouri Nitin P. Badhe |
author_facet | Girish N. Swamy Conal Quah Elmunzar Bagouri Nitin P. Badhe |
author_sort | Girish N. Swamy |
collection | DOAJ |
description | This paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered fully and was back to his regular activities. He presented with a history of sudden onset pain and locking of the left knee since the preceding three months. There was no history of trauma, and the patient was mobilizing with difficulty using crutches. Radiographs revealed fracture of the posterior condyle of the femoral prosthesis. Revision surgery was performed as an elective procedure revealing the broken prosthesis. The TC3RP-PFC revision prosthesis was used with a medial parapatellar approach. The patient recovered fully without any squeal. Mechanical failure of the knee arthroplasty prosthesis is rare, and nontraumatic fracture of the femoral metallic component has not been reported before. |
format | Article |
id | doaj-art-63642182ff3640d1bdbed6603cf85124 |
institution | Kabale University |
issn | 2090-6749 2090-6757 |
language | English |
publishDate | 2014-01-01 |
publisher | Wiley |
record_format | Article |
series | Case Reports in Orthopedics |
spelling | doaj-art-63642182ff3640d1bdbed6603cf851242025-02-03T01:24:00ZengWileyCase Reports in Orthopedics2090-67492090-67572014-01-01201410.1155/2014/896348896348Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical FailureGirish N. Swamy0Conal Quah1Elmunzar Bagouri2Nitin P. Badhe3Trauma and Orthopaedics, Queens Medical Centre, Nottingham University Hospitals NHS Trust, Nottingham, UKTrauma and Orthopaedics, Queens Medical Centre, Nottingham University Hospitals NHS Trust, Nottingham, UKTrauma and Orthopaedics, Queens Medical Centre, Nottingham University Hospitals NHS Trust, Nottingham, UKTrauma and Orthopaedics, Queens Medical Centre, Nottingham University Hospitals NHS Trust, Nottingham, UKThis paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered fully and was back to his regular activities. He presented with a history of sudden onset pain and locking of the left knee since the preceding three months. There was no history of trauma, and the patient was mobilizing with difficulty using crutches. Radiographs revealed fracture of the posterior condyle of the femoral prosthesis. Revision surgery was performed as an elective procedure revealing the broken prosthesis. The TC3RP-PFC revision prosthesis was used with a medial parapatellar approach. The patient recovered fully without any squeal. Mechanical failure of the knee arthroplasty prosthesis is rare, and nontraumatic fracture of the femoral metallic component has not been reported before.http://dx.doi.org/10.1155/2014/896348 |
spellingShingle | Girish N. Swamy Conal Quah Elmunzar Bagouri Nitin P. Badhe Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure Case Reports in Orthopedics |
title | Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure |
title_full | Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure |
title_fullStr | Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure |
title_full_unstemmed | Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure |
title_short | Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure |
title_sort | nontraumatic fracture of the femoral condylar prosthesis in a total knee arthroplasty leading to mechanical failure |
url | http://dx.doi.org/10.1155/2014/896348 |
work_keys_str_mv | AT girishnswamy nontraumaticfractureofthefemoralcondylarprosthesisinatotalkneearthroplastyleadingtomechanicalfailure AT conalquah nontraumaticfractureofthefemoralcondylarprosthesisinatotalkneearthroplastyleadingtomechanicalfailure AT elmunzarbagouri nontraumaticfractureofthefemoralcondylarprosthesisinatotalkneearthroplastyleadingtomechanicalfailure AT nitinpbadhe nontraumaticfractureofthefemoralcondylarprosthesisinatotalkneearthroplastyleadingtomechanicalfailure |