Delayed Aneurysm Rupture Following Endovascular Treatment with Contour Device: A Case Report

Delayed rupture of intracranial aneurysms after endovascular treatment is a rare but serious complication. We report the first documented case of late aneurysmal rupture following treatment with a Contour intrasaccular device. A patient in their 60s with a basilar tip aneurysm underwent endovascular...

Full description

Saved in:
Bibliographic Details
Main Authors: Osman Koc, Mostafa Mahmoud, Ehab Mahmoud, Ali Ayyad, Ahmad Own
Format: Article
Language:English
Published: Korean Society of Interventional Neuroradiology 2025-03-01
Series:Neurointervention
Subjects:
Online Access:http://neurointervention.org/upload/pdf/neuroint-2024-00493.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Delayed rupture of intracranial aneurysms after endovascular treatment is a rare but serious complication. We report the first documented case of late aneurysmal rupture following treatment with a Contour intrasaccular device. A patient in their 60s with a basilar tip aneurysm underwent endovascular treatment using a 14-mm Contour device. Fifteen months later, the patient presented with a fatal intraventricular hemorrhage, and imaging revealed device displacement and aneurysm growth. This case underscores the importance of meticulous device sizing and follow-up, especially for large aneurysms.