Delayed Aneurysm Rupture Following Endovascular Treatment with Contour Device: A Case Report
Delayed rupture of intracranial aneurysms after endovascular treatment is a rare but serious complication. We report the first documented case of late aneurysmal rupture following treatment with a Contour intrasaccular device. A patient in their 60s with a basilar tip aneurysm underwent endovascular...
Saved in:
| Main Authors: | , , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Korean Society of Interventional Neuroradiology
2025-03-01
|
| Series: | Neurointervention |
| Subjects: | |
| Online Access: | http://neurointervention.org/upload/pdf/neuroint-2024-00493.pdf |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | Delayed rupture of intracranial aneurysms after endovascular treatment is a rare but serious complication. We report the first documented case of late aneurysmal rupture following treatment with a Contour intrasaccular device. A patient in their 60s with a basilar tip aneurysm underwent endovascular treatment using a 14-mm Contour device. Fifteen months later, the patient presented with a fatal intraventricular hemorrhage, and imaging revealed device displacement and aneurysm growth. This case underscores the importance of meticulous device sizing and follow-up, especially for large aneurysms. |
|---|---|