Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta
The use of telescopic intramedullary rods for the treatment of lower extremity deformity in children with osteogenesis imperfecta has been well described. The reinforcement of these weakened weightbearing bones with intramedullary devices leads to improvements in mobility that progress into adulthoo...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2021-01-01
|
Series: | Case Reports in Pediatrics |
Online Access: | http://dx.doi.org/10.1155/2021/9982289 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832547637123874816 |
---|---|
author | Michael E. Kahan Nathan R. Angerett Jill C. Flanagan |
author_facet | Michael E. Kahan Nathan R. Angerett Jill C. Flanagan |
author_sort | Michael E. Kahan |
collection | DOAJ |
description | The use of telescopic intramedullary rods for the treatment of lower extremity deformity in children with osteogenesis imperfecta has been well described. The reinforcement of these weakened weightbearing bones with intramedullary devices leads to improvements in mobility that progress into adulthood. Although the current telescopic intramedullary rod systems are an upgrade from earlier systems, they are still associated with high rates of reoperation and complication. We describe a unique complication encountered during a revision which involved the male retriever system for the Fassier–Duval rod (FDR) (Pega Medical, Quebec, CA) experienced intraoperatively. To our knowledge, this mechanism of failure has not been previously described in the literature. |
format | Article |
id | doaj-art-599f6485b94b43609dc64d5fe53d57e7 |
institution | Kabale University |
issn | 2090-6803 2090-6811 |
language | English |
publishDate | 2021-01-01 |
publisher | Wiley |
record_format | Article |
series | Case Reports in Pediatrics |
spelling | doaj-art-599f6485b94b43609dc64d5fe53d57e72025-02-03T06:43:57ZengWileyCase Reports in Pediatrics2090-68032090-68112021-01-01202110.1155/2021/99822899982289Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis ImperfectaMichael E. Kahan0Nathan R. Angerett1Jill C. Flanagan2Department of Orthopaedic Surgery, UPMC Pinnacle, Harrisburg, PA, USADepartment of Orthopaedic Surgery, UPMC Pinnacle, Harrisburg, PA, USADepartment of Orthopaedics and Sports Medicine, Children’s Hospital of Atlanta, Atlanta, GA, USAThe use of telescopic intramedullary rods for the treatment of lower extremity deformity in children with osteogenesis imperfecta has been well described. The reinforcement of these weakened weightbearing bones with intramedullary devices leads to improvements in mobility that progress into adulthood. Although the current telescopic intramedullary rod systems are an upgrade from earlier systems, they are still associated with high rates of reoperation and complication. We describe a unique complication encountered during a revision which involved the male retriever system for the Fassier–Duval rod (FDR) (Pega Medical, Quebec, CA) experienced intraoperatively. To our knowledge, this mechanism of failure has not been previously described in the literature.http://dx.doi.org/10.1155/2021/9982289 |
spellingShingle | Michael E. Kahan Nathan R. Angerett Jill C. Flanagan Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta Case Reports in Pediatrics |
title | Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta |
title_full | Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta |
title_fullStr | Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta |
title_full_unstemmed | Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta |
title_short | Intraoperative Hardware Failure of the Fassier–Duval Rescue System in a Pediatric Patient with Osteogenesis Imperfecta |
title_sort | intraoperative hardware failure of the fassier duval rescue system in a pediatric patient with osteogenesis imperfecta |
url | http://dx.doi.org/10.1155/2021/9982289 |
work_keys_str_mv | AT michaelekahan intraoperativehardwarefailureofthefassierduvalrescuesysteminapediatricpatientwithosteogenesisimperfecta AT nathanrangerett intraoperativehardwarefailureofthefassierduvalrescuesysteminapediatricpatientwithosteogenesisimperfecta AT jillcflanagan intraoperativehardwarefailureofthefassierduvalrescuesysteminapediatricpatientwithosteogenesisimperfecta |