Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms

Computed Tomography (CT) has been one of the most used exam for radiologic diagnostic in medicine. The increase of CT is a global concern due to high doses of radiation. The dose evaluation in CT is one of many steps that can contribute for reducing patient doses. The head CT scans helps to diagnose...

Full description

Saved in:
Bibliographic Details
Main Author: Fernanda Stephanie Santos
Format: Article
Language:English
Published: Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR) 2020-09-01
Series:Brazilian Journal of Radiation Sciences
Subjects:
Online Access:https://bjrs.org.br/revista/index.php/REVISTA/article/view/936
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850104299799969792
author Fernanda Stephanie Santos
author_facet Fernanda Stephanie Santos
author_sort Fernanda Stephanie Santos
collection DOAJ
description Computed Tomography (CT) has been one of the most used exam for radiologic diagnostic in medicine. The increase of CT is a global concern due to high doses of radiation. The dose evaluation in CT is one of many steps that can contribute for reducing patient doses. The head CT scans helps to diagnose disorders that affect the brain, including tumors, infarction, bleeding within the brain, hematoma and other diseases. The aim of this work is to compare the reduction of absorbed dose in the head CT scan with and without the use of bismuth shielding using a male and female anthropomorphic phantom model Alderson Rando.The head CT scan were done from the cervical vertebra C1 to the top of skull, in a GE CT scanner, LightSpeed VCT model, with 64 channels.Radiochromic films strips were used to evaluate the doses in the organs such as lenses, thyroid, hypophysis, spinal cord, pharynx, breasts, salivary and parotid glands.Record doses were lower with the use of bismuth shielding in both phantoms for all organs, mainly in the lenses. The analysis of noise in the image of the head central slice presented acceptable values for soft tissues, less than 1%. Dose values were significantly reduced and they suggested that the use of bismuth shielding would be a proper procedure for protection during a head CT scan.
format Article
id doaj-art-575da37007674f7e9fb4ebdc224686d8
institution DOAJ
issn 2319-0612
language English
publishDate 2020-09-01
publisher Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR)
record_format Article
series Brazilian Journal of Radiation Sciences
spelling doaj-art-575da37007674f7e9fb4ebdc224686d82025-08-20T02:39:21ZengBrazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR)Brazilian Journal of Radiation Sciences2319-06122020-09-0181B (Suppl.)10.15392/bjrs.v8i1B.936731Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using PhantomsFernanda Stephanie Santos0Federal University of Minas GeraisComputed Tomography (CT) has been one of the most used exam for radiologic diagnostic in medicine. The increase of CT is a global concern due to high doses of radiation. The dose evaluation in CT is one of many steps that can contribute for reducing patient doses. The head CT scans helps to diagnose disorders that affect the brain, including tumors, infarction, bleeding within the brain, hematoma and other diseases. The aim of this work is to compare the reduction of absorbed dose in the head CT scan with and without the use of bismuth shielding using a male and female anthropomorphic phantom model Alderson Rando.The head CT scan were done from the cervical vertebra C1 to the top of skull, in a GE CT scanner, LightSpeed VCT model, with 64 channels.Radiochromic films strips were used to evaluate the doses in the organs such as lenses, thyroid, hypophysis, spinal cord, pharynx, breasts, salivary and parotid glands.Record doses were lower with the use of bismuth shielding in both phantoms for all organs, mainly in the lenses. The analysis of noise in the image of the head central slice presented acceptable values for soft tissues, less than 1%. Dose values were significantly reduced and they suggested that the use of bismuth shielding would be a proper procedure for protection during a head CT scan.https://bjrs.org.br/revista/index.php/REVISTA/article/view/936computed tomographybismuth shieldingdosimetry
spellingShingle Fernanda Stephanie Santos
Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
Brazilian Journal of Radiation Sciences
computed tomography
bismuth shielding
dosimetry
title Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
title_full Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
title_fullStr Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
title_full_unstemmed Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
title_short Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
title_sort comparative study of the use of bismuth shielding on head computed tomography scans using phantoms
topic computed tomography
bismuth shielding
dosimetry
url https://bjrs.org.br/revista/index.php/REVISTA/article/view/936
work_keys_str_mv AT fernandastephaniesantos comparativestudyoftheuseofbismuthshieldingonheadcomputedtomographyscansusingphantoms