Development and validation of a RP-HPLC method for the quantisation studies of metronidazole and furazolidone from product Enteroguard M

An isocratic high-performance liquid chromatography (HPLC) procedure was developed for quantitative determination of metronidazole and furazolidone in tablet dosage forms of ENTEROGUARD M. HPLC separation was carried out by reversed phase chromatography Kromasil C18 (250mm x 4,6mm i.d.; 5μm particle...

Full description

Saved in:
Bibliographic Details
Main Authors: Elena Gabriela Oltean, Milea, B.
Format: Article
Language:English
Published: Asociaţia Naţională a Fabricanţilor de Produse de Uz Veterinar 2010-12-01
Series:Medicamentul Veterinar
Subjects:
Online Access:http://www.veterinarypharmacon.com/docs/610-2010-%20ART.7.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An isocratic high-performance liquid chromatography (HPLC) procedure was developed for quantitative determination of metronidazole and furazolidone in tablet dosage forms of ENTEROGUARD M. HPLC separation was carried out by reversed phase chromatography Kromasil C18 (250mm x 4,6mm i.d.; 5μm particle size), held at 25°C. The mobile phase consisted of methanol/ 0,1% phosphoric acid aq. (20/80v/v), run at flow rate of 1 mL/ min and with UV detection at 317nm. Method validation investigated parameters such as linearity (r2 = 0.9999), range, precision, accuracy and specificity. The described method can be successfully applied for the analysis of ENTEROGUARD M tablets.
ISSN:1843-9527