Fourth-generation Ceramic Head Fracture in Total Hip Arthroplasty: A Case Report and Literature Review
Ceramics are used in total hip arthroplasty due to inherent wettability and low wear rates, but fracture risk is a known complication. Rates as high as 13.4% were reported in the past, yet as low as 0.02% in newer generations. Howard et al. reported a fracture rate of 0.009% for fourth-generation ce...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2025-04-01
|
Series: | Arthroplasty Today |
Subjects: | |
Online Access: | http://www.sciencedirect.com/science/article/pii/S2352344125000019 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Ceramics are used in total hip arthroplasty due to inherent wettability and low wear rates, but fracture risk is a known complication. Rates as high as 13.4% were reported in the past, yet as low as 0.02% in newer generations. Howard et al. reported a fracture rate of 0.009% for fourth-generation ceramic heads. We present a case report of a 69-year-old male with a BMI of 40.01 kg/m2 who suffered a fracture of a 36-mm ceramic femoral head with pseudoacetabular involvement of the polyethylene liner. This occurred 2 years and 3 months after his primary surgery after a 4-foot fall off of a ladder. The patient underwent revision of his total hip arthroplasty with arthroplasty of the polyethylene liner and femoral head component without stem explantation. |
---|---|