Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible

Objective To compare the effect of different scanning parameters of cone beam computed tomography (CBCT) on displaying trabecular microstructure in the anterior region of the mental foramen of the mandible, and to provide a basis for the rational selection of CBCT scanning parameters. Methods This s...

Full description

Saved in:
Bibliographic Details
Main Author: DONG Qi, FENG Yongjing, GAO Antian, LIN Zitong
Format: Article
Language:zho
Published: Editorial Department of Journal of Prevention and Treatment for Stomatological Diseases 2024-12-01
Series:口腔疾病防治
Subjects:
Online Access:https://www.kqjbfz.com/fileup/2096-1456/PDF/2096-1456(2024)12-0945-09.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850115054222966784
author DONG Qi, FENG Yongjing, GAO Antian, LIN Zitong
author_facet DONG Qi, FENG Yongjing, GAO Antian, LIN Zitong
author_sort DONG Qi, FENG Yongjing, GAO Antian, LIN Zitong
collection DOAJ
description Objective To compare the effect of different scanning parameters of cone beam computed tomography (CBCT) on displaying trabecular microstructure in the anterior region of the mental foramen of the mandible, and to provide a basis for the rational selection of CBCT scanning parameters. Methods This study was approved by the Ethics Committee of the Affiliated Stomatological Hospital, Medical School of Nanjing University. An in vitro study was conducted using CBCT (ProMax 3D Mid) to scan eight dry human mandibular specimens with five scanning protocols: Group A: 90 kV/6.3 mA, Group B: 90 kV/8.0 mA, Group C: 90 kV/10.0 mA, Group D: 75 kV/8.0 mA, and Group E: 60 kV/8.0 mA, resulting in a total of 40 CBCT images. Entrance surface dose (ESD) and computed tomography dose index (CTDI) under different scanning conditions were recorded. The original CBCT images were imported into the image analysis software (Hiscan Analyzer) to measure four trabecular bone microstructural parameters in the region of interest of the mandible: trabecular thickness (Tb.Th), trabecular number (Tb.N), trabecular space (Tb.Sp), and bone volume/tissue volume (BV/TV). A total of 108 CBCT images were retrospectively collected from clinical implant patients using any of the 90 kV/6.3 mA, 90 kV/8.0 mA, or 90 kV/10.0 mA scanning conditions, and the above four parameters of the region of interest of the mandible were also measured. SPSS 26.0 software was used to compare the differences in the four trabecular bone microstructural parameters between the CBCT images of the mandibular specimens in vitro and clinical patients in vivo. Results The in vitro study results showed that reducing the tube voltage and tube current would lead to a decrease in the ESD and CTDI. When the tube voltage was maintained at 90 kV and the tube current was changed, BV/TV, Tb.N, and Tb.Th values increased with the increase of tube current; Tb.Sp values decreased with the increase of tube current, but there was no statistically significant difference in the four bone morphological parameters (<i>P </i>&gt; 0.05). When the tube current was maintained at 8.0 mA and the tube voltage was altered, BV/TV and Tb.N decreased with the increase of tube voltage, Tb.Sp values increased with the increase of tube voltage, and BV/TV, Tb.N, and Tb.Sp showed statistically significant differences (<i>P</i> &lt; 0.05). In clinical patients undergoing CBCT scanning, when the tube voltage was 90 kV and the tube current was different (6.3, 8.0, 10.0 mA), there was no statistically significant difference in the four bone morphological parameters (<i>P </i>&gt; 0.05). Conclusions In this study, when the tube voltage was fixed at 90 kV, there was no difference in the trabecular microstructure of the anterior region of the mandible when the tube current was increased. When CBCT scanning of clinical patients needs to show the trabecular microstructure of the anterior region of the mandible, the tube current can be appropriately reduced to decrease the radiation dose received by the patient. Thus, it is recommended to use the parameters 90 kV and 6.3 mA for CBCT scanning.
format Article
id doaj-art-4ecc24f12f184f5c8f5aa4dca211c419
institution OA Journals
issn 2096-1456
language zho
publishDate 2024-12-01
publisher Editorial Department of Journal of Prevention and Treatment for Stomatological Diseases
record_format Article
series 口腔疾病防治
spelling doaj-art-4ecc24f12f184f5c8f5aa4dca211c4192025-08-20T02:36:41ZzhoEditorial Department of Journal of Prevention and Treatment for Stomatological Diseases口腔疾病防治2096-14562024-12-01321294595310.12016/j.issn.2096-1456.202440303Effects of different scanning parameters of cone beam CT on displaying microstructure of mandibleDONG Qi, FENG Yongjing, GAO Antian, LIN Zitong0Department of Dentomaxillofacial Radiology, Nanjing Stomatological Hospital, Affiliated Hospital of Medical School, Institute of Stomatology, Nanjing University, Nanjing 210008, ChinaObjective To compare the effect of different scanning parameters of cone beam computed tomography (CBCT) on displaying trabecular microstructure in the anterior region of the mental foramen of the mandible, and to provide a basis for the rational selection of CBCT scanning parameters. Methods This study was approved by the Ethics Committee of the Affiliated Stomatological Hospital, Medical School of Nanjing University. An in vitro study was conducted using CBCT (ProMax 3D Mid) to scan eight dry human mandibular specimens with five scanning protocols: Group A: 90 kV/6.3 mA, Group B: 90 kV/8.0 mA, Group C: 90 kV/10.0 mA, Group D: 75 kV/8.0 mA, and Group E: 60 kV/8.0 mA, resulting in a total of 40 CBCT images. Entrance surface dose (ESD) and computed tomography dose index (CTDI) under different scanning conditions were recorded. The original CBCT images were imported into the image analysis software (Hiscan Analyzer) to measure four trabecular bone microstructural parameters in the region of interest of the mandible: trabecular thickness (Tb.Th), trabecular number (Tb.N), trabecular space (Tb.Sp), and bone volume/tissue volume (BV/TV). A total of 108 CBCT images were retrospectively collected from clinical implant patients using any of the 90 kV/6.3 mA, 90 kV/8.0 mA, or 90 kV/10.0 mA scanning conditions, and the above four parameters of the region of interest of the mandible were also measured. SPSS 26.0 software was used to compare the differences in the four trabecular bone microstructural parameters between the CBCT images of the mandibular specimens in vitro and clinical patients in vivo. Results The in vitro study results showed that reducing the tube voltage and tube current would lead to a decrease in the ESD and CTDI. When the tube voltage was maintained at 90 kV and the tube current was changed, BV/TV, Tb.N, and Tb.Th values increased with the increase of tube current; Tb.Sp values decreased with the increase of tube current, but there was no statistically significant difference in the four bone morphological parameters (<i>P </i>&gt; 0.05). When the tube current was maintained at 8.0 mA and the tube voltage was altered, BV/TV and Tb.N decreased with the increase of tube voltage, Tb.Sp values increased with the increase of tube voltage, and BV/TV, Tb.N, and Tb.Sp showed statistically significant differences (<i>P</i> &lt; 0.05). In clinical patients undergoing CBCT scanning, when the tube voltage was 90 kV and the tube current was different (6.3, 8.0, 10.0 mA), there was no statistically significant difference in the four bone morphological parameters (<i>P </i>&gt; 0.05). Conclusions In this study, when the tube voltage was fixed at 90 kV, there was no difference in the trabecular microstructure of the anterior region of the mandible when the tube current was increased. When CBCT scanning of clinical patients needs to show the trabecular microstructure of the anterior region of the mandible, the tube current can be appropriately reduced to decrease the radiation dose received by the patient. Thus, it is recommended to use the parameters 90 kV and 6.3 mA for CBCT scanning.https://www.kqjbfz.com/fileup/2096-1456/PDF/2096-1456(2024)12-0945-09.pdfcone-beam computed tomography|tube voltage|tube current|trabecular microstructure|trabecular bone microarchitectural parameters|trabecular thickness|trabecular number|trabecular space|bone volume /tissue volume
spellingShingle DONG Qi, FENG Yongjing, GAO Antian, LIN Zitong
Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible
口腔疾病防治
cone-beam computed tomography|tube voltage|tube current|trabecular microstructure|trabecular bone microarchitectural parameters|trabecular thickness|trabecular number|trabecular space|bone volume /tissue volume
title Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible
title_full Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible
title_fullStr Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible
title_full_unstemmed Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible
title_short Effects of different scanning parameters of cone beam CT on displaying microstructure of mandible
title_sort effects of different scanning parameters of cone beam ct on displaying microstructure of mandible
topic cone-beam computed tomography|tube voltage|tube current|trabecular microstructure|trabecular bone microarchitectural parameters|trabecular thickness|trabecular number|trabecular space|bone volume /tissue volume
url https://www.kqjbfz.com/fileup/2096-1456/PDF/2096-1456(2024)12-0945-09.pdf
work_keys_str_mv AT dongqifengyongjinggaoantianlinzitong effectsofdifferentscanningparametersofconebeamctondisplayingmicrostructureofmandible