Influence of abutment material on peri-implant clinical parameters: a prospective study

Objective: This study evaluated the peri-implant clinical parameters of zirconia versus metallic abutments for standardized custom-made implantsupported single-tooth restorations. Material and Methods: Sixteen subjects received 30 dental implants on esthetic region of upper jaws. Test group had 14...

Full description

Saved in:
Bibliographic Details
Main Authors: Fabiano Cortez Zanardo, José Della Pasqua Neto, Luiz Henrique Alves Madeira, Jamil Awad Shibli, Keize Távora Barbosa, José Augusto Rodrigues, Alessandra Cassoni
Format: Article
Language:English
Published: Universidade Estadual Paulista 2014-08-01
Series:Brazilian Dental Science
Online Access:https://ojs.ict.unesp.br/index.php/cob/article/view/972
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849228647112638464
author Fabiano Cortez Zanardo
José Della Pasqua Neto
Luiz Henrique Alves Madeira
Jamil Awad Shibli
Keize Távora Barbosa
José Augusto Rodrigues
Alessandra Cassoni
author_facet Fabiano Cortez Zanardo
José Della Pasqua Neto
Luiz Henrique Alves Madeira
Jamil Awad Shibli
Keize Távora Barbosa
José Augusto Rodrigues
Alessandra Cassoni
author_sort Fabiano Cortez Zanardo
collection DOAJ
description Objective: This study evaluated the peri-implant clinical parameters of zirconia versus metallic abutments for standardized custom-made implantsupported single-tooth restorations. Material and Methods: Sixteen subjects received 30 dental implants on esthetic region of upper jaws. Test group had 14 implant-supported single-tooth restorations with ceramic abutment and control group had 16 restorations with metallic abutment. Peri-implant clinical parameters were recorded after 110 ± 10 days in function at six sites by each implant, in each subject. The means of clinical measures: probing depth (PD) and clinical attachment level (CAL); mean of percentages of sites with visible plaque (VP), bleeding on probing (BOP) and suppuration (SU) were recorded to each subject. To verify differences between groups Mann Whitney U-test was selected PD and CAL and k-light test VP, BOP and SU. Results: One zirconia implant-supported single-tooth restoration fractured at insertion time. The mean of PD, CAL, and BOP of the samples sites were lower to test group than control (p < 0.05). There were no differences of mean levels of VP and S between test and control group (p > 0.05). Conclusion: Clinical parameters around implant-supported restorations were influenced by abutment type with better results to zirconia abutments, at least, after short time period evaluation. 
format Article
id doaj-art-475c82a0c77148d7a2aae79677c9aeef
institution Kabale University
issn 2178-6011
language English
publishDate 2014-08-01
publisher Universidade Estadual Paulista
record_format Article
series Brazilian Dental Science
spelling doaj-art-475c82a0c77148d7a2aae79677c9aeef2025-08-22T17:54:40ZengUniversidade Estadual PaulistaBrazilian Dental Science2178-60112014-08-0117310.14295/bds.2014.v17i3.972Influence of abutment material on peri-implant clinical parameters: a prospective studyFabiano Cortez Zanardo0José Della Pasqua Neto1Luiz Henrique Alves Madeira2Jamil Awad Shibli3Keize Távora Barbosa4José Augusto Rodrigues5Alessandra Cassoni6Dental Research Division - Department of Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil.Dental Research Division - Department of Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil.Dental Research Division - Department of Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil.Dental Research Division - Department of Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil.Dental Research Division - Department of Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil.Dental Research Division - Department of Restorative and Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil.Dental Research Division - Department of Restorative and Implant Dentistry - Guarulhos University - Guarulhos - SP - Brazil. Objective: This study evaluated the peri-implant clinical parameters of zirconia versus metallic abutments for standardized custom-made implantsupported single-tooth restorations. Material and Methods: Sixteen subjects received 30 dental implants on esthetic region of upper jaws. Test group had 14 implant-supported single-tooth restorations with ceramic abutment and control group had 16 restorations with metallic abutment. Peri-implant clinical parameters were recorded after 110 ± 10 days in function at six sites by each implant, in each subject. The means of clinical measures: probing depth (PD) and clinical attachment level (CAL); mean of percentages of sites with visible plaque (VP), bleeding on probing (BOP) and suppuration (SU) were recorded to each subject. To verify differences between groups Mann Whitney U-test was selected PD and CAL and k-light test VP, BOP and SU. Results: One zirconia implant-supported single-tooth restoration fractured at insertion time. The mean of PD, CAL, and BOP of the samples sites were lower to test group than control (p < 0.05). There were no differences of mean levels of VP and S between test and control group (p > 0.05). Conclusion: Clinical parameters around implant-supported restorations were influenced by abutment type with better results to zirconia abutments, at least, after short time period evaluation.  https://ojs.ict.unesp.br/index.php/cob/article/view/972
spellingShingle Fabiano Cortez Zanardo
José Della Pasqua Neto
Luiz Henrique Alves Madeira
Jamil Awad Shibli
Keize Távora Barbosa
José Augusto Rodrigues
Alessandra Cassoni
Influence of abutment material on peri-implant clinical parameters: a prospective study
Brazilian Dental Science
title Influence of abutment material on peri-implant clinical parameters: a prospective study
title_full Influence of abutment material on peri-implant clinical parameters: a prospective study
title_fullStr Influence of abutment material on peri-implant clinical parameters: a prospective study
title_full_unstemmed Influence of abutment material on peri-implant clinical parameters: a prospective study
title_short Influence of abutment material on peri-implant clinical parameters: a prospective study
title_sort influence of abutment material on peri implant clinical parameters a prospective study
url https://ojs.ict.unesp.br/index.php/cob/article/view/972
work_keys_str_mv AT fabianocortezzanardo influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy
AT josedellapasquaneto influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy
AT luizhenriquealvesmadeira influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy
AT jamilawadshibli influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy
AT keizetavorabarbosa influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy
AT joseaugustorodrigues influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy
AT alessandracassoni influenceofabutmentmaterialonperiimplantclinicalparametersaprospectivestudy